Journal of Clinical and Experimental Ophthalmology

Journal of Clinical and Experimental Ophthalmology
Open Access

ISSN: 2155-9570

Research Article - (2016) Volume 7, Issue 3

A Comparison of Three Techniques for Measurement of Intraocular Pressure in Normal Eyes

Haya M. Al Farhan*
Department of Optometry and Vision Sciences, College of Applied Medicine Sciences, King Saud University, Kingdom of Saudi Arabia
*Corresponding Author: Haya M. Al Farhan, Department of Optometry and Vision Sciences, College of Applied Medicine Sciences, King Saud University, P.O. Box 10219, Riyadh 11433, Kingdom of Saudi Arabia, Tel: + 966 508381107 Email:

Abstract

Objectives: To compare the precision of intraocular pressure (IOP) measurements acquired using an ocular response analyser (ORA), Auto Kerato-Refracto-Tonometer (TRK-1P), and Goldmann applanation tonometer (GAT) in healthy eyes.
Methods: In this prospective study, one eye of each of 57 normal subjects was randomly selected for analysis. Measurements of the IOP were performed using ORA, TRK-1P, and GAT, and measurements of corneal hysteresis (CH), corneal resistance factor (CRF), and central corneal thickness (CCT) were performed using ORA. Repeatability was assessed by the coefficient of variation (CV) and interclass correlation coefficients (ICCs). Agreement among tonometers was assessed by Bland–Altman plots and one-way ANOVA.
Results: The average IOPs measured using Goldmann-correlated IOP (IOPg), corneal-compensated IOP (IOPcc), TRK-1P, and GAT (± SDs) were 15.13 ± 2.76, 14.39 ± 2.59, 16.54 ± 2.93, and 15.21 ± 2.54 mmHg, respectively. Intra-observer agreement across all tonometers was strong and slightly higher for GAT and IOPg than for TRK-1P. The intra-observer CVs for GAT IOPg, and TRK-1P, were 4.22 (ICC=0.94), 4.99 (ICC=0.93), and 6.69 (ICC=0.86), respectively. Inter-observer agreement between various measurement methods was evaluated with Bland-Altman plots with multiple measurements per subject and ICCs. Results indicated fairly poor agreement across measurement methods, as supported by large limits of agreement and ICCs.
Conclusion: GAT, ORA, and TRK-1P are highly reliable methods for measurement of the IOP; however, the instruments cannot be used interchangeably.

Keywords: Goldmann applanation tonometer; TRK-1P; Ocular response analyser; Central corneal thickness; Corneal resistance factor; Corneal hysteresis; Normal eyes

Introduction

Glaucoma is the second leading cause of blindness worldwide. Quigley and Broman (2006) estimated that by 2020, 1,618,718 individuals in the Middle East would have glaucoma; the corresponding estimate for 2020 is 2,295,407 individuals [1]. The intraocular pressure (IOP) measurement is vital for monitoring the effectiveness of treatment and evaluating the risk of glaucoma progression. Since IOP reduction is the objective of treatment, accurate IOP assessment is essential for monitoring the efficacy of therapy and for assessing the risk of glaucomatous progression [2,3].

The Goldmann applanation tonometer (GAT) has been considered the gold standard for measuring IOP for a number of decades [3]. However, GAT measurement of IOP is associated with calibration errors [4,5], contamination [6,7], and certain corneal features, such as central corneal thickness (CCT) [8,9], corneal curvature [10], and axial length [10]. Recently, advances in technology, including the dynamic contour tonometer (DCT; PASCAL, Ziemer Ophthalmic System, Port, Switzerland), ocular response analyser (ORA; Reichert Ophthalmic Instruments, Buffalo, NY, USA), and Auto Kerato-Refracto-Tonometer (TRK-1P); Topcon Corporation, Tokyo, Japan), have led to the introduction of several instruments to increase the accuracy of IOP measurements and reduce the risk of infection [11]. However, it is unclear which of these instruments is best for precise measurement of IOP.

ORA analyses four variables: the Goldmann-correlated IOP (IOPg), corneal-compensated IOP (IOPcc), corneal hysteresis (CH), and corneal resistance factor (CRF) [3-6]. CH is an indication of viscous damping in the cornea; IOPcc is an IOP measurement that is less affected by corneal properties than GAT; and CRF is a measurement of the cumulative effects of both the viscous and elastic resistance encountered by the air jet while deforming the corneal surface. Various studies have compared GAT with DCT and ORA [11]. However, no comparisons have been made between GAT or ORA and TRK-1P.

Therefore, the aim of this study was to assess the intra-observer repeatability and agreement of three tonometry instruments: ORA, TRK-1P, and GAT. To the best of our knowledge, this first study to compare TRK-1P with GAT and ORA.

Subjects and Methods

Fifty-seven eyes from 57 healthy, oculovisually normal subjects (32 women and 25 men) were enrolled in this prospective study. The subjects were randomly selected from a clinical population. The mean age (± standard deviation [SD]) was 24.50 ± 5.85 years (range, 19-40 years). Comprehensive anterior and posterior segment examinations of all subjects were performed using a slit lamp and direct ophthalmoscope. The exclusion criteria included contact lens wearers, positive history (or objective signs) of ocular disease, and systemic disease with ocular implications, such diabetes mellitus. CCT measurements are affected by age and refractive error [12]. Thus, the spherical equivalent refractive error was ± 4.00 D or more, the corneal astigmatism was -3.00 D or greater, and the corneal curvature was 48 D or greater. The corneal curvature was determined by an autorefractometer (Auto Kerato-Refracto-Tonometer [TRK-1P]; Topcon Corporation, Tokyo, Japan). The CCT measurements were obtained using Visante OCT from Zeiss (Carl Zeiss Meditec, Dublin, CA, USA). To eliminate the disadvantages of ultrasound pachymetry, such as physical contact with the cornea, topical anesthesia was used; the accuracy of measurements was dependent on the perpendicularity of the probe’s application to the cornea, and reproducibility relied on precise probe placement on the corneal centre [13].

For each subject, one eye was randomly selected using a table of random numbers generated in Microsoft Excel (Microsoft Corporation, Redmond, WA, USA). An unmasked single investigator conducted all measurements of the IOP, CCT, CRF, and CH. All IOP, CCT, CRF, and CH measurements were collected between 10:00 and 14:00 h [14].

The Auto Kerato-Refracto Tonometer (TRK-1P) is a new instrument ideal for screening that combines four measurements: autorefractometry, keratometry, noncontact tonometry, and pachymetry. Using rotary prism technology, the TRK-1P provides unmatched accuracy and reliability. The innovative optical design incorporated into the TRK-1P allows for accurate and reliable measurements with a pupil as small as f2 mm. The TRK-1P is easy to operate. A significant advantage of noncontact tonometry is the elimination of possible risks related to all contact tonometry, such as corneal abrasion, use of topical anaesthetic or fluorescein, and spread of infection.

Three consecutive measurements were performed using each method for each subject. However, three subjects dropped out of the study as they were apprehensive about being examined with GAT; these subjects were excluded. The purpose of the study was explained to all subjects, and informed consent was obtained from each subject before beginning the examination. The study was conducted in accordance with the ethical principles described in the 2008 Declaration of Helsinki, and the study protocol was approved by the research ethics review board of the College of Applied Medicine Science at King Saud University.

Three IOP instruments

The first measurements were conducted with noncontact tonometers followed by GAT to eliminate the possible effects of applanation on the hysteresis value. The IOPg, IOPcc, CRF, and CH measurements were obtained using ORA, followed by measurement of IOP with TRK-1P. All ORA measurements were obtained at wave scores 6 and above [15]. The mean of the three readings from each tonometer was recorded. Then, CCT measurements were obtained using Visante OCT. There was a resting interval of around 15 min between noncontact tonometry and GAT. Finally, the IOP was measured with GAT after instillation of one drop of 0.5% proparacaine in the lower conjunctival sac, along with one drop of fluorescein sodium (0.25%). Three readings were obtained from GAT, and the mean was recorded.

Statistical methods

Demographic data for all subjects and repeatability were assessed by determining the coefficient of variation (CV), interclass correlation coefficient (ICC), and 95% confident intervals (CIs). Bland-Altman analysis was performed to determine agreement between methods of measurement with multiple observations per individual, and the average of agreement between methods was used to assess the limits of agreement using Medcalc software version 11.4.4.0.

Results

The study included 57 normal subjects (34 right eyes and 23 left eyes). The mean spherical equivalent of refractive error was -0.36 ± 0.67 D. Data for GAT, TRK-1P, IOPg, IOPcc, CH, CRF, and CCT are summarised in Table 1.

Parameters Mean ± SD**
Age (years) 24.50 ± 5.85
GAT* 15.21 ± 2.54
TRK-1P* 16.54 ± 2.93
IOPg* 15.13 ± 2.76
IOPcc* 14.39 ± 2.59
CRF* 11.10 ± 1.80
CH* 11.40 ± 1.40
CCT (µm)            546.43 ± 31.54

Table 1: Characteristics of normal individuals.

Intra-observer repeatability of IOP measurements

Intra-observer repeatability of IOP measurements obtained with the three instruments is summarised in Table 2. The results indicated acceptable agreement between measurements of the GAT and IOPg, as shown by the semisymmetrical distribution of the Bland-Altman plot and an ICC in the high agreement range (ICC=0.941). The remainder of measurement combinations did not yield sufficient agreement, as shown by the widely distributed Bland-Altman plots and ICCs below the critical value of 0.80.

Comparisons Limits of agreement CV% ICC 95% CI†
GAT vs. IOPg -2.40 to 2.60* 4.24 0.941 0.902 to 0.965
GAT vs. IOPcc -2.80 to 4.40* 8.01 0.796 0.590 to 0.892
GAT vs. TRK-1P -5.10 to 2.10* 9.35 0.729 0.167 to .0890
TRK-1P vs. IOPg -1.70 to 4.80* 9.04 0.768 0.081 to 0.918
TRK-1P vs. IOPcc -2.80 to 7.40* 14.65 0.473 0-.015 to 0.734
*mmHg, †confident intervals

Table 2: Measurements of the mean differences ± SDs and limits of agreement between the three instruments.

Comparisons of the TRK-1P and the IOPcc yielded the poorest agreement. Overall, these results suggested that while each tonometer method was by itself reliable and consistent, scores obtained using different clinical methods may not be directly comparable (Figures 1 and 2).

clinical-experimental-ophthalmology-Bland-Altman plots

Figure 1: Bland-Altman plots of agreement showing the mean differences and limits of agreement of IOP measurements with multiple observations per subject. (A) GAT versus IOPg, (B) GAT versus IOPcc, and (C) GAT versus TRK-IP.

clinical-experimental-ophthalmology-IOP-measurements

Figure 2: Bland-Altman plots of agreement showing the mean differences and limits of agreement of IOP measurements with multiple observations per subject. (A) TRK-IP vs. IOPg, and (B) TRK-IP vs. IOPcc.

Agreement between instruments

Consistency across repeated trials within each method was examined using CVs and Cronbach’s alpha intra-class correlations (Table 3). Across all tonometers, consistency and reliability was in the excellent range (as>0.90; ICCs>0.90). GAT and IOPg had slightly better consistency compared with TRK-1P; however, these differences were not meaningfully different. These results indicated that measurements within a single tonometer instrument were consistent and reliable.

 Repeatability
Instruments GAT IOPg   TRK-1P
CV% 4.22 4.99 6.69
ICC 0.94 0.93 0.86
95% CI† 0.905 to 0.961 0.890 to 0.954 0.789 to 0.9012
†Confident interval

Table 3: Intra-observer repeatability of IOP measurements with the three instruments.

Bland-Altman plots of the mean differences and limits of agreement (LOAs) of IOP measurements using GAT versus IOPg, IOPcc, and TRK-IP and using TRK-IP versus IOPg and IOPcc are shown in Figure 3.

clinical-experimental-ophthalmology-GAT-versus-IOPg

Figure 3: Bland-Altman plot of agreement showing the mean differences and limits of agreement of IOP measurements using GAT versus IOPg, IOPcc, and TRK-IP and using TRK-IP versus IOPg and IOPcc.

Discussion

Precise, reliable assessment of IOP measurements is important for maintaining ocular visual function, decision-making regarding treatment modalities in patients with glaucoma, and improvements in treatment regimens [16].

Intra-observer repeatability of IOP measurements obtained with the three instruments demonstrated that GAT and IOP had higher levels of reliability and repeatability and good agreement for multiple observations per individual. The ICCs of GAT and IOPcc, GAT and TRK-IP, and TRK-IP and IOPg showed good levels of reliability and repeatability below 0.80, yet did not yield sufficient agreement for multiple observations per individual, as shown by the wide distributions of the Bland-Altman. In contrast, the ICC TRK-IP and IOPcc measurements yielded poor levels of reliability, repeatability, and agreement of multiple observations per individual. In general, the results suggested that measurements of IOP obtained using different techniques may not be directly comparable.

Furthermore, the results showed that GAT and IOPg had higher levels of reliability and repeatability (as indicated by CVs and ICCs) than TRK-IP. Statistical analysis revealed that there was high intraobserver repeatability for GAT and IOPg, but high intra-observer variability for TRK-1P. Bland-Altman analyses of IOP measurements showed poor LOAs between GAT and TRK-IP, TRK-IP and IOPg, and TRK-IP and IOPcc, as shown by the wide distributions of the Bland- Altman plots. However, good agreement was noted between GAT and IOPg and between GAT and IOPcc, with mean differences of 0.1 and 0.81 mmHg, respectively. Overall, these results suggested that each tonometer was in and of itself reliable and consistent.

Several of studies have reported high intra-observer repeatability of IOP measurements obtained using GAT and IOPg in normal eyes [17-19]. Our results show high intra-observer consistency and repeatability of IOP measurements obtained with GAT and IOPg. However, Wang et al. [20] reported higher CVs (9.70 and 7.0, respectively) and ICCs (0.79 and 0.79, respectively) for GAT and IOPg. This discrepancy could be due to the sample size used in their study. No previous reports have examined the relative performance of TRKIP. This study demonstrated that the TRK-IP instrument showed good intra-observer consistency and repeatability of IOP measurements.

In this study, the 95% limits of agreement of the between GAT and IOPg, GAT and TRK-IP, TRK-IP and IOPg, and TRK-IP and IOPcc were outside the accepted limits, indicating that they were not interchangeable. For GAT and the noncontact tonometer, this may be because noncontact tonometry tended to overestimate the GAT at high IOPs and underestimate GAT at low IOPs due to the CCT effect [21-23]. However, comparisons between noncontact tonometers could be explained by the use of different calculation methods to obtain the IOP [22].

Conclusion

GAT, ORA, and TRK-1P are highly reliable for measurement of the IOP; however, the instruments cannot be used interchangeably.

References

  1. Quigley HA, Broman AT (2006) The number of people with glaucoma worldwide in 2010 and 2020. Br J Ophthalmol 90: 262-267.
  2. Leske MC, Heijl A, Hussein M, Bengtsson B, Hyman L, et al. (2003) Factors for glaucoma progression and the effect of treatment: the early manifest glaucoma trial. Arch Ophthalmol 121: 48-56.
  3. [No authors listed] (2000) The Advanced Glaucoma Intervention Study (AGIS): 7. The relationship between control of intraocular pressure and visual field deterioration.The AGIS Investigators. Am J Ophthalmol 130: 429-440.
  4. ElMallah MK, Asrani SG (2008) New ways to measure intraocular pressure. CurrOpinOphthalmol 19: 122-126.
  5. Medeiros FA, Weinreb RN (2006) Evaluation of the influence of corneal biomechanical properties on intraocular pressure measurements using the ocular response analyzer. J Glaucoma 15: 364-370.
  6. Martinez-de-la-Casa JM, Garcia-Feijoo J, Fernandez-Vidal A, Mendez-Hernandez C, Garcia-Sanchez J (2006) Ocular response analyzer versus Goldmannapplanation tonometry for intraocular pressure measurements. Invest Ophthalmol Vis Sci 47: 4410-4414.
  7. Kotecha A, Elsheikh A, Roberts CR, Zhu H, Garway-Heath DF (2006) Corneal thickness- and age-related biomechanical properties of the cornea measured with the ocular response analyzer. Invest Ophthalmol Vis Sci 47: 5337-5347.
  8. Broman AT, Congdon NG, Bandeen-Roche K, Quigley HA (2007) Influence of corneal structure, corneal responsiveness, and other ocular parameters on tonometric measurement of intraocular pressure. J Glaucoma 16: 581-588.
  9. Brubaker RF (1999) Tonometry and corneal thickness. Arch Ophthalmol 117: 104-105.
  10. Ehlers N, Bramsen T, Sperling S (1975) Applanation tonometry and central corneal thickness. ActaOphthalmol (Copenh) 53: 34-43.
  11. Renier C, Zeyen T, Fieuws S, Vandenbroeck S, Stalmans I (2010) Comparison of ocular response analyzer, dynamic contour tonometer and Goldmannapplanation tonometer. IntOphthalmol 30: 651-659.
  12. Tonnu PA, Ho T, Newson T, El Sheikh A, Sharma K, et al. (2005) The influence of central corneal thickness and age on intraocular pressure measured by pneumotonometry, non-contact tonometry, the Tono-Pen XL, and Goldmannapplanation tonometry. Br J Ophthalmol 89: 851-854.
  13. Piñero DP, Plaza AB, Alió JL (2008) Anterior segment biometry with 2 imaging technologies: very-high-frequency ultrasound scanning versus optical coherence tomography. J Cataract Refract Surg 34: 95-102.
  14. Kotecha A, Crabb DP, Spratt A, Garway-Heath DF (2009) The relationship between diurnal variations in intraocular pressure measurements and central corneal thickness and corneal hysteresis. Invest Ophthalmol Vis Sci 50: 4229-4236.
  15. Mandalos A, Anastasopoulos E, Makris L, Dervenis N, Kilintzis V, et al. (2013) Inter-examiner reproducibility of Ocular Response Analyzer using the waveform score quality index in healthy subjects. J Glaucoma 22: 152-155.
  16. Nakamoto K, Yasuda N, Fukuda T (2008) [Correlation of age and intraocular pressure with visual field damage in patients with normal-tension glaucoma]. Nippon GankaGakkaiZasshi 112: 371-375.
  17. Sullivan-Mee M, Gerhardt G, Halverson KD, Qualls C (2009) Repeatability and reproducibility for intraocular pressure measurement by dynamic contour, ocular response analyzer, and goldmannapplanation tonometry. J Glaucoma 18: 666-673.
  18. Moreno-Montañés J, Maldonado MJ, García N, Mendiluce L, García-Gómez PJ, et al. (2008) Reproducibility and clinical relevance of the ocular response analyzer in nonoperated eyes: corneal biomechanical and tonometric implications. Invest Ophthalmol Vis Sci 49: 968-974.
  19. Ouyang PB, Li CY, Zhu XH, Duan XC (2012) Assessment of intraocular pressure measured by Reichert Ocular Response Analyzer, GoldmannApplanation Tonometry, and Dynamic Contour Tonometry in healthy individuals. Int J Ophthalmol 5: 102-107.
  20. Wang AS, Alencar LM, Weinreb RN, Tafreshi A, Deokule S, et al. (2013) Repeatability and reproducibility of Goldmannapplanation, dynamic contour, and ocular response analyzer tonometry. J Glaucoma 22: 127-132.
  21. Hong J, Xu J, Wei A, Deng SX, Cui X, et al. (2013) A new tonometer--the Corvis ST tonometer: clinical comparison with noncontact and Goldmannapplanationtonometers. Invest Ophthalmol Vis Sci 54: 659-665.
  22. Tonnu PA, Ho T, Sharma K, White E, Bunce C, et al. (2005) A comparison of four methods of tonometry: method agreement and interobserver variability. Br J Ophthalmol 89: 847-850.
  23. Kretz G, Demailly P (1992) X-PERT NCT advanced logic tonometer valuation. IntOphthalmol 16: 287-290.
Citation: Al Farhan HM (2016) A Comparison of Three Techniques for Measurement of Intraocular Pressure in Normal Eyes. J Clin Exp Ophthalmol 7:553.

Copyright: © 2016 Al Farhan HM. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Top