Accuracy of Bedside Upper Airway Ultrasonography vs. Standard Auscultation for Assuring the Location of Endotracheal Tube after Tracheal Intubation: Comparative Study Controlled by Quantitative Waveform Capnography

Sherif Arafa, Amir Abouzkry* and Aymen Mohamady
Department of Anesthesia, ICU and Pain Management, Faculty of Medicine, Aswan University, Egypt
*Corresponding Author: Amir Abouzkry, Department of Anesthesia, ICU and Pain Management, Faculty of Medicine, Aswan University, Egypt, Tel: 00201090969655 Email:

Abstract

Objectives: To estimate the accuracy and rapidity of bed-side upper airway ultrasonography (UA-US) versus standard auscultation (SA) for confirmation of endotracheal tube (ETT) position.
Patients and methods: 107 patients underwent tracheal intubation for maintenance of general anesthesia. Position of ETT was confirmed by chest auscultation for the presence of breath sounds on both sides, and by UAUS using 9–12 MHz linear US transducer that was placed transversely on the neck anteriorly and superior to the suprasternal notch to visualize the ETT in the tracheal transverse and longitudinal views. The data obtained were compared to findings on using waveform capnography (WC). Time to define ETT position was determined. Study outcomes included determination of ETT position; tracheal or esophageal, accuracy of diagnosis and time taken till confirming the diagnosis.
Results: With comparison to WC findings, UA-US revealed sensitivity, specificity and accuracy rates of 97%, 71.4% and 95.3% while that for SA were 93.6%, 53.9% and 88.8%, respectively with significantly higher specificity and accuracy rates for UA-WC versus SA. Time required for confirmation of ETT position was significantly shorter with WC (9.16 ± 0.69 sec.) compared to SA and UA-US with significant difference in favor of UA-US (11.14 ± 1.3 vs. 13.5 ± 2.15 sec).
Conclusion: Confirmation of ETT position using UA-US or WC is very important because of the high false result depending on SA alone. UA-US using bed-side equipment is a simple, accurate and fast method than SA compared to WC as a gold standard, so it is recommended to be one of the important theater equipments whenever possible

Keywords: Endotracheal tube position; Ultrasonography; Capnography; Standard auscultation

Introduction

Mastering skill of endotracheal intubation to secure airway plays an important role in many settings [1] and confirmation of correct endotracheal tube (ETT) position immediately after intubation for general anesthesia has been one of the most challenging issues of airway management [2].

Confirmation of correct placement of ETT should be done rapidly [3] for early detection of esophageal intubation through a reliable method [4] as tube malposition is associated with various serious adverse outcomes [3].

Multiple methods were used over decades for verification of ETT position, end tidal carbon dioxide was found to be the most appropriate for identifying esophageal intubation [5]. However, there are reports assuring the greater accuracy of the esophageal detector device especially in emergency settings [6]. Capnography was documented to be the best reliable method for fast detection of ETT position [7].

Clinicians utilize auscultation of breath sounds to verify correct ETT placement. However, anesthesia providers often delay timely charting of ET intubation and this latency may result in decreased efficacy of clinical decision support systems [8]. Bedside ultrasonography had been widely used for preliminary diagnoses especially in bedridden patients [9] and in emergency medicine [10] and to aid for diagnostic decision making in emergency department [11].

Aim of Work

The current study aimed to assess accuracy and rapidity of bed-side upper airway ultrasonography (UA-US) versus standard auscultation (SA) for confirmation of endotracheal tube (ETT) position after tracheal intubation.

• Setting: University Hospital, Aswan, Egypt

• Design: Comparative double-blinded study

Patients And Methods

The current study was conducted since Oct 2016 till Oct 2017. The study protocol was approved by the Local Ethical Committee and all enrolled patients signed written fully informed consent to participate in the study.

Inclusion criteria included adult patients of ASA grade I and II and were assigned for general anesthesia with endotracheal intubation. Exclusion criteria included presence of severe neck trauma, neck mass, scar of previous neck surgery, anatomic neck distortion, limited neck extension, abnormal airway anatomy and obesity. Patients need emergency intubation or with high risk of aspiration and severe cardiac, pulmonary, hepatic, or renal disease were excluded from the study.

Patients were examined the night before surgery for determination of demographic data, clinical assessment for assurance of inclusion and exclusion criteria. Preoperatively, baseline hemodynamic data including heart rate (HR), blood pressure measures; systolic (SBP), diastolic (DBP) and mean arterial pressure (MAP) were non-invasively determined. Also, arterial peripheral oxygen saturation was determined using pulse oximetry.

All patients received the same anesthetic protocol including premedication using intravenous (IV) midazolam 2 mg, induction of anesthesia by fentanyl 2 μg/kg and propofol 2 mg/kg followed by atracuraium 0.5 mg/kg to facilitate tracheal intubation. Patients were maintained on controlled mask ventilation with 100% O2 until adequately relaxed and then endotracheal tube was inserted within 3-5 min.

Induction of anesthesia and intubation were applied by an anesthetist who was blinded about the results of testing the accuracy of tube position. This anesthetist was responsible for auscultatory confirmation of ETT position by the presence of breath sounds sought on both sides by auscultation on the infra-clavicular and infra-axillary areas and were classified into definite, doubtful or absent.

Time taken for auscultatory confirmation was defined as the time elapsed from beginning to the end of auscultation. Also, this anesthetist was responsible for recording blood pressure and heart rate readings and O2 saturation immediately, 5 min and 15 min after intubation.

Another anesthetist was responsible for performing upper airway US (UA-US) to ascertain the position of ETT; tracheal or esophageal and time elapsed since probe positioning till giving a decision about the tube position was given is determined. UA-US was performed using SonoScape US machine (China).

A 9-12 MHz linear ultrasound transducer was placed transversely on the neck anteriorly and superior to the suprasternal notch before intubation. Immediately after intubation, the transducer probe was used to visualize the ETT in the tracheal longitudinal and transverse views. The probe was then moved to the left to look at the esophagus to assure whether it was empty or distended by the tube.

The obtained results were compared to the positive result of quantitative waveform capnography (WC) using Dräger Fabius® plus XL (Germany). WC was considered as the gold standard for confirmation of ETT position.

A positive result on WC was determined by a 3rd anesthetist who was blinded about the results of SA and UA-US and was set as the detection of appropriate exhaled CO2 that must be >30 mmHg after at least 5 breaths, with a typical CO2 waveform. Time used by WC to assure tracheal intubation was also determined.

Study outcomes

• The detection of ETT position; tracheal or esophageal

• The accuracy of diagnosis using either SA standard auscultatory method or UA-US in comparison to WC.

• Time taken till confirming the diagnosis

Statistical analysis

Sample size was calculated using the standard nomogram proposed by Kraemer and Thiemann [12] and a sample size of ≥ 100 patients was determined to be sufficient to detect a difference at 5% significance level and give the trial 80% power [13].

Obtained data were presented as mean ± SD, ranges, numbers and percentages. Results were analyzed paired t-test and Chi-square test (X2 test). Test diagnostic validity characters; namely sensitivity, specificity and accuracy were calculated. Statistical analysis was conducted using the IBM SPSS (Version 23, 2015) for Windows statistical package. P value <0.05 was considered statistically significant.

Results

The study included 126 patients; 19 patients were excluded and 107 patients were enrolled in the study (Figure 1). Enrollment data of studied patients are shown in Table 1.

anesthesia-clinical-research-Flow-Chart

Figure 1: Flow Chart of the Study.

Data Findings
Age (years) <30 years Frequency 19 (17.8%)
Mean (± SD) 23.9 ± 2.1
30-40 Frequency 30 (28%)
Mean (± SD) 35.5 ± 2.3
>40-50 Frequency 26 (24.3%)
Mean (± SD) 45.6 ± 2
>50 Frequency 32 (29.9%)
Mean (± SD) 53.5 ± 1.4
Total Mean (± SD) 41.4 ± 10.9
Gender Male 68 (63.6%)
Female  39 (36.4%)
Body mass index data Weight (kg) 75.1 ± 6.7
Height (cm) 170 ± 3.3
BMI (kg/m2) <25 52 (48.6%)
>25 55 (51.4%)
Mean (± SD) 26 ± 2.2
ASA grade I  86 (80.4%)
II  21 (19.6%)
Data are presented as mean ± SD & numbers; percentages are in parenthesis; BMI: Body Mass Index; p>0.05: Indicates non-significant difference.

Table 1: Patients' enrollment data.

All patients showed significantly increased HR and blood pressure measures immediately and 5 min after intubation compared to baseline measures. Fifteen minutes after heart rate and blood pressure measures were significantly lower compared to baseline measures (Table 2).

Data Findings
HR (beats/min) Baseline 84.6 ± 5.6
Immediate after intubation 89.3 ± 6.7*
5 min after intubation 92.2 ± 5.6*
15 min after intubation 80.4 ± 8.4*
SBP (mmHg) Baseline 115.6 ± 8.5
Immediate after intubation 127.1 ± 8.9*
5 min after intubation 97.6 ± 9.1*
15 min after intubation 103.8 ± 7.2*
DBP (mmHg) Baseline 75.5 ± 3
Immediate after intubation 77.2 ± 3.5*
5 min after intubation 79.6 ± 5*
15 min after intubation 76.5 ± 3.9*
MAP (mmHg) Baseline 88.9 ± 3.3
Immediate after intubation 93.8 ± 3.4*
5 min after intubation 85.6 ± 4.7*
15 min after intubation 86.2 ± 3.6*
SaO2 (%) Baseline 97.5 ± 1.6
Immediate after intubation 96.4 ± 3*
5 min after intubation 97.9 ± 1.2*
15 min after intubation 98.3 ± 0.5*
Data are presented as mean ± SD; HR: Heart rate; SBP: Systolic Blood Pressure; DBP: Diastolic Blood Pressure; MAP: Mean Arterial Pressure; SaO2: Arterial Oxygen Saturation; *: indicates significant difference vs. baseline value.

Table 2: Patients' hemodynamic and arterial oxygen saturation data determined immediately, 5 min and 15 min after intubation compared to preoperative findings.

Auscultatory confirmation of ETT position assured proper tracheal intubation in 94 patients (87.9%) with definitely sought breath sounds, but suspected aberrant ETT position in 13 patients (12.1%) as doubtfully sought breath sounds in 10 patients and absent breath sounds in three patients.

On the other hand, UA-US detected tracheal position of ETT (Figure 2) in 99 patients (92.5%), and esophageal position of ETT (Figure 3) in 8 patients (7.5%). Considering WC, as a gold standard for comparison, defined proper tracheal position of ETT (Figure 4) in 100 patients (93.5%) and esophageal position of ETT (Figure 5) in 7 patients (6.5%), (Table 3).

Value Auscultatory UA-US WC
Positive Total 94 (87.9%) 99 (92.5%) 100 (93.5%)
True Positive 88 (93.6%) 97 (98%)
False Positive 6 (6.4%) 2 (2%)
Negative Total 13 (12.1%) 8 (7.5%) 7 (6.5%)
True Negative 7 (53.8%) 5 (62.5%)
False Negative 6 (46.2%) 3 (37.5%)
Data are presented as numbers; percentages are in parenthesis; UA-US: Upper Airway Ultrasonography; WC: Waveform Capnography.

Table 3: Evaluation of diagnostic findings by auscultatory and UA-US confirmation of ETT position versus WC confirmation.

anesthesia-clinical-research-horizontal-view

Figure 2: A) UA-US tracheal horizontal view showing two hyperchoic lines which mean that ETT is inside the trachea; B) UAUS tracheal vertical view showing two symmetrical hyperchoic lines which mean that ETT is inside the trachea.

anesthesia-clinical-research-hyperchoic

Figure 3: UA-US tracheal horizontal view showing two hyperchoic lines inside the esophagus which appeared dilated on the left side (Yellow Arrow) with absence of intratracheal lines (Red Arrow).

anesthesia-clinical-research-Shows-positive

Figure 4: Shows positive end-tidal CO2 quantitative WC which means correct tracheal placement of ETT.

anesthesia-clinical-research-shows-negative

Figure 5: shows negative end-tidal CO2 quantitative WC which means esophageal placement of ETT.

Evaluation of diagnostic validity of auscultatory confirmation of ETT position versus findings of WC showed that auscultatory confirmation showed sensitivity and specificity rates of 93.6% and 53.9%, respectively with an accuracy rate for defining proper ETT position of 88.8%.

On contrary, UA-US confirmation of proper ETT position versus WC findings showed sensitivity and specificity rates of 97% and 71.4%, respectively with an accuracy rate of 95.3% with significantly higher specificity and accuracy rates versus auscultatory confirmation (Figure 6).

anesthesia-clinical-research-Diagnostic-validity

Figure 6: Diagnostic validity characters of auscultatory vs. UA-US confirmation of ETT position vs. WC.

Furthermore, UA-US confirmation of ETT position allowed significantly (p<0.001) faster detection of ETT position than auscultatory confirmation (11.14 ± 1.3 vs. 13.5 ± 2.15 s), despite both modalities consumed significantly (p<0.001) longer time till confirmation of ETT position in comparison to time required using WC (9.16 ± 0.69 s) for detection of ETT (Figure 7).

anesthesia-clinical-research-Mean-time

Figure 7: Mean time take till confirmation of ETT position.

Discussion

The current study tried to evaluate the accuracy of upper airway ultrasonography (UA-US) for detection of endotracheal tube (ETT) position; tracheal or esophageal in comparison to auscultatory confirmation. Quantitative waveform capnography (WC) was used as a gold standard for comparison. In line with the reliance on WC, Langhan et al. [14] documented that WC use had significantly increased overtime and its use during intubation was not associated with esophageal intubation or occurrence of cardiac arrest. Also, Scoccimarro et al. [15] reported that WC improves the time to intubation and first pass success rates through more consistent and expedient means of detecting optimal intubating conditions. Moreover, Silvestri et al. [16] approved the reliability of WC for verifying ETT location during low-perfusion states such as cardiac arrest, and suggested that WC showed no false negatives or positives, thus indicating 100% sensitivity and specificity.

In comparison to WC findings, auscultatory confirmation of ETT position showed sensitivity and specificity rates of 93.6% and 53.9%, respectively with an accuracy rate for defining proper ETT position of 88.8%, while UA-US showed sensitivity and specificity rates of 97% and 71.4%, respectively with an accuracy rate of 95.3% for confirmation of proper ETT position, with significant difference versus auscultatory method.

In line with the reported figures for diagnostic validity of UA-US for confirmation of ETT position, Alonso Quintela et al. [9] reported that US appears to be as effective as capnography for identifying ETT and may be useful wherever capnography is less reliable as during cardiopulmonary resuscitation. Thereafter, Das et al. [17] reported that transtracheal US is a useful tool to confirm endotracheal intubation with sensitivity and specificity rates of 98% in both emergency and elective intubation. Also, Tessaro et al. [18] assured that US visualization of a saline-inflated ETT cuff at the suprasternal notch is accurate and rapid modality to confirm the insertion depth of ETT in children with sensitivity of 98.8% and specificity of 96.4%.

Recently, Masoumi et al. [4] found WC detected appropriate tracheal placement in 100% of patients, while in comparison US detected tracheal intubation in 93% and esophageal intubation in 7% with sensitivity, specificity, and positive and negative predictive values for detecting appropriate ETT tracheal placement of 98.9%, 100%, 100% and 85.7%, respectively. Moreover, Lahham et al. [19] performed real-time transverse tracheal sonography during intubation to evaluate correct ETT placement and detected correct tracheal placement in 94.4% of patients, while 5.6% of intubations were interpreted as esophageal, and concluded that US conferred sensitivity and specificity rates of 98.5% and 75%, respectively.

The reported superiority of UA-US over auscultatory confirmation of ETT position goes in hand with Ramsingh et al. [20] reported that sensitivity and specificity rates of US for differentiating tracheal versus bronchial intubations were 93% and 96% vs. 66% and 59% for auscultation and the accuracy for identification of tracheal versus bronchial intubation was 95% versus 62% with significant difference in favor of US. Also, Zamani Moghadam et al. [21] evaluated the diagnostic accuracy of US in ETT placement confirmation compared to a combination of 4 clinical confirmation methods and found tracheal US had excellent sensitivity (>90%) and good specificity (80-90%) for confirming ETT placement. Moreover, Abhishek et al. [2] documented that both capnography and upper airway USG may be used as primary procedures for the confirmation of ETT placement.

As another support for the superiority of US over auscultatory confirmation of ETT position, Álvarez-Díaz et al. [22] compared trans-thoracic lung US and clinical method in confirming the position of double-lumen tube in thoracic anesthesia and found the sensitivity and specificity of US was 98.6% and 52.9%, while that of the clinical method was 84.5% and 41.1% with significant difference in favor of US. Recently, Rajan et al. [23] assessed the efficacy of US sliding lung sign versus standard auscultation for confirming ETT placement in normal, overweight and obese surgical patients and concluded that US confirmation of ETT placement is superior in accuracy compared to auscultatory method.

Furthermore, UA-US allowed significantly faster detection of ETT position than auscultatory confirmation, despite of the significantly longer time consumed by both modalities compared to time required using WC. In line with these findings, Pfeiffer et al. [24,25] documented that verification of ETT placement with US is as fast as auscultation alone and faster than the combination of auscultation and capnography, even in obese patients [23]. Thereafter, Alonso Quintela et al. [9] reported despite US was slower than capnography for identifying ETT, US is as effective and quicker than X-ray for assessment of ETT insertion depth, and it may contribute to decrease the routine use of X-ray after tracheal intubation. Recently, Rajan et al. [23] documented that US confirmation of ETT placement is superior in speed compared to auscultatory method. Also, Thomas et al. [26] reported that US confirmed ETT placement with comparable sensitivity and specificity to quantitative WC and clinical methods, but it yielded results considerably faster than the other two modalities.

Conclusion

It can be concluded that confirmation of ETT position using either UA-US or WC is a mandatory requirement because of the high false diagnosis depending on standard auscultatory confirmation alone. Using bed-side UA-US is an easy, accurate and fast method than standard auscultation compared to WC as a gold standard, so it is suggested to be one of the essential theater equipments whenever possible.

References

  1. Ghatehorde NK, Regunath H (2017) Intubation, Endotracheal Tube, Medications. StatPearls [Internet]. Treasure Island (FL): StatPearls Publishing.
  2. Abhishek C, Munta K, Rao SM, Chandrasekhar CN (2017) End-tidal capnography and upper airway ultrasonography in the rapid confirmation of endotracheal tube placement in patients requiring intubation for general anaesthesia. Indian J Anaesth 61: 486-489.
  3. Sharma D, Ahmad Tabatabaii S, Farahbakhsh N (2017) Role of ultrasound in confirmation of endotracheal tube in neonates: A review. J Matern Fetal Neonatal Med 1-9.
  4. Masoumi B, Azizkhani R, Emam GH, Asgarzadeh M, Kharazi BZ (2017) Predictive Value of Tracheal Rapid Ultrasound Exam Performed in the Emergency Department for Verification of Tracheal Intubation. Open Access Maced J Med Sci 5: 618-623.
  5. Kugelman A, Zeiger-Aginsky D, Bader D, Shoris I, Riskin A (2008) A novel method of distal end-tidal CO2 capnography in intubated infants: comparison with arterial CO2 and with proximal mainstream end-tidal CO2. Pediatrics 122: e1219-1224.
  6. Kalmar AF, Absalom A, Monsieurs KG (2012) A novel method to detect accidental esophageal intubation based on ventilation pressure waveforms. Resuscitation 83: 177-182.
  7. Walsh BK, Crotwell DN, Restrepo RD (2011) Capnography/Capnometry during mechanical ventilation: 2011. Respir Care 56: 503-509.
  8. Jalali A, Rehman M, Lingappan A, Nataraj C (2016) Automatic detection of endotracheal intubation during the anesthesia procedure. J Dyn Syst Meas Control 138: 1110131-11101318.
  9. Alonso Quintela P, Oulego Erroz I, Mora Matilla M, Rodríguez Blanco S, Mata Zubillaga D, et al. (2014) Usefulness of bedside ultrasound compared to capnography and X-ray for tracheal intubation. An Pediatr (Barc) 81: 283-288.
  10. Vieira RL, Levy JA (2010) Bedside ultrasonography to identify hip effusions in pediatric patients. Ann Emerg Med 55: 284-289.
  11. Summers SM, Scruggs W, Menchine MD, Lahham S, Anderson C, et al. (2010) A prospective evaluation of emergency department bedside ultrasonography for the detection of acute cholecystitis. Ann Emerg Med 56: 114-122.
  12. Kraemer HC, Theimann S (1987) How many subjects? Statistical power analysis in research; Newbury Park, CA: Sage.
  13. Murphy KR, Myors B (2003) Statistical Power Analysis: A Simple and General Model for Traditional and Modern Hypothesis Tests; (2nd Edn)., Lawrence Erlbaum Associates, Inc.
  14. Langhan ML, Emerson BL, Nett S, Pinto M, Harwayne-Gidansky I, et al.  Pediatric Acute Lung Injury and Sepsis Investigators (PALISI) and National Emergency Airway Registry for Children (NEAR4KIDS) Investigators. End-Tidal Carbon Dioxide Use for Tracheal Intubation: Analysis from the National Emergency Airway Registry for Children (NEAR4KIDS) Registry.
  15. Scoccimarro A, West JR, Kanter M, Caputo ND (2017) Waveform capnography: an alternative to physician gestalt in determining optimal intubating conditions after administration of paralytic agents. Emerg Med J 35: 62-64.
  16. Silvestri S, Ladde JG, Brown JF, Roa JV, Hunter C, et al. (2017) Endotracheal tube placement confirmation: 100% sensitivity and specificity with sustained four-phase capnographic waveforms in a cadaveric experimental model. Resuscitation 115: 192-198.
  17. Das SK, Choupoo NS, Haldar R, Lahkar A (2015) Transtracheal ultrasound for verification of endotracheal tube placement: a systematic review and meta-analysis. Can J Anaesth 62: 413-423.
  18. Tessaro MO, Salant EP, Arroyo AC, Haines LE, Dickman E (2015) Tracheal rapid ultrasound saline test (T.R.U.S.T.) for confirming correct endotracheal tube depth in children. Resuscitation 89: 8-12.
  19. Lahham S, Baydoun J, Bailey J, Sandoval S, Wilson SP, et al. (2017) A Prospective Evaluation of Transverse Tracheal Sonography during Emergent Intubation by Emergency Medicine Resident Physicians. J Ultrasound Med 36: 2079-2085.
  20. Ramsingh D, Frank E, Haughton R, Schilling J, Gimenez KM, et al. (2016) Auscultation versus Point-of-care Ultrasound to Determine Endotracheal versus Bronchial Intubation: A Diagnostic Accuracy Study. Anesthesiology 124: 1012-1020.
  21. Zamani Moghadam H, Sharifi MD, Rajabi H, Mousavi Bazaz M, Alamdaran A, et al. (2017) Screening Characteristics of Bedside Ultrasonography in Confirming Endotracheal Tube Placement; a Diagnostic Accuracy Study. Emerg (Tehran) 5: e19.
  22. Álvarez-Díaz N, Amador-García I, Fuentes-Hernández M, Dorta-Guerra R (2015) Comparison between transthoracic lung ultrasound and a clinical method in confirming the position of double-lumen tube in thoracic anesthesia. A pilot study. Rev Esp Anestesiol Reanim 62: 305-312.
  23. Rajan S, Surendran J, Paul J, Kumar L (2017) Rapidity and efficacy of ultrasonographic sliding lung sign and auscultation in confirming endotracheal intubation in overweight and obese patients. Indian J Anaesth 61: 230-234.
  24. Pfeiffer P, Rudolph SS, Børglum J, Isbye DL (2011) Temporal comparison of ultrasound vs. auscultation and capnography in verification of endotracheal tube placement. Acta Anaesthesiol Scand 55: 1190-1195.
  25. Pfeiffer P, Bache S, Isbye DL, Rudolph SS, Rovsing L, et al. (2012) Verification of endotracheal intubation in obese patients - temporal comparison of ultrasound vs. auscultation and capnography. Acta Anaesthesiol Scand 56: 571-576.
  26. Thomas VK, Paul C, Rajeev PC, Palatty BU (2017) Reliability of Ultrasonography in Confirming Endotracheal Tube Placement in an Emergency Setting. Indian J Crit Care Med 21: 257-261.
Citation: Arafa S, Abouzkry A, Mohamady A (2018) Accuracy of Bedside Upper Airway Ultrasonography vs. Standard Auscultation for Assuring the Location of Endotracheal Tube after Tracheal Intubation: Comparative Study Controlled by Quantitative Waveform Capnography. J Anesth Clin Res 9: 804.

Copyright: © 2018 Arafa S, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Top