Journal of Antivirals & Antiretrovirals

Journal of Antivirals & Antiretrovirals
Open Access

ISSN: 1948-5964

+44 1300 500008

Research Article - (2021)Volume 13, Issue 4

Concordance of "Rapid" Serological Tests and IgG and IgM Chemiluminescence for SARS-COV-2

Sáenz-Flor K1,2* and Santafé Lorena M2
 
*Correspondence: Sáenz-Flor K, School of Medicine, Central University, Quito, Ecuador, Email:

Author info »

Abstract

Objectives: To evaluate a sample of immunofluorescent and immunochromatographic rapid tests used in Ecuador to show their agreement in relation to Chemiluminescence.

Setting: Primary care limitations for a sample of “rapid test” assays used for serological diagnosis of SARS-CoV-2 in Ecuador, South-America.

Participants: A diagnostic test evaluation assay was performed to establish the performance of five “rapid” tests for IgG and IgM serology for SARS-CoV-2 using a panel of 30 serum samples from routine patients.

Interventions: For the evaluation of clinical performance, the qualitative results of the "rapid" tests were compared against those obtained by chemiluminescence, dichotomized as positives (>10 AU/mL) or negative (<10 UA/mL).

Primary and secondary outcome measures: Demonstration of agreement in defining the subjects with the dichotomous criterion (positive or negative for SARS-CoV-2 antibodies), calculating complex repeatability, positive agreements and negative agreements, with their corresponding 95% confidence interval and Cohen's Kappa test.

Results: The best agreement is seen in the immunofluorescent assay, for the IgG contrast, with a particularly good kappa index (0.85), without positive disagreements and a negative disagreement of about 15%. In the immunochromatographic methods Kappa index was 0.61 at best, with disagreements in negative findings of ≈35% and in positive cases of up to ≈70%.

Conclusions: Given the high demand and supply in the market of "rapid serological tests", its evaluation against panels of serologically positive or negative samples established by Chemiluminescence or Electro chemiluminescence is essential to authorize its extensive use in populations.

Keywords

COVID-19; Chemiluminescence; Rapid test

Introduction

In early 2020, the World Health Organization declared a global health emergency due to the outbreak of a new coronavirus, initially called 2019-nCOV and later called SARS-CoV-2, causing a severe acute respiratory syndrome, Corona Virus Disease 2019 (COVID-19) [1,2].

It is a zoonotic coronavirus that causes, in its severe forms, an acute respiratory distress syndrome that in most cases can occur with mild symptoms [3,4]. As of May 12, an overall case fatality of 6.9% (283,153 deaths/4,088,848 confirmed cases) is reported, in the Americas 5.9% (104,549 deaths/1,774,371 confirmed cases), and for Ecuador 7.3% (2,145 deaths/29509 confirmed cases) according to the official bulletins of the World Health Organization [5].

The COVID-19 serological tests for IgG and IgM have been developed with several methodologies: Immunoenzymatic Assay (ELISA), Chemiluminescence, Electro Chemiluminescence, Fluorescent Lateral Flow Immunoassays and Immunochromatography. Each one has a different and variable performance in relation to the clinical moment in which they are used, as well as whether they were developed to detect S (Spike) or N (Nucleocapsid) antigens, being the former apparently more sensitive [6]. None of these tests should be used for the diagnosis or population screening of the disease, considering that the antibodies appear only on the 8th-14th day of the disease onset [7].

Laboratory tests that detect antibodies to SARS-CoV-2, including rapid immunodiagnostic tests, need validation to determine their accuracy and reliability, since there is a double risk. The first one is to falsely label people who have been infected as sero-negative, and the second one is that people who have not been infected are falsely labeled as sero-positive; both errors have serious consequences. In addition to this, it is necessary to ensure that these tests distinguish between infections caused by SARS-CoV-2 and those caused by other human coronaviruses (cross reaction) [8].

The rapid development associated with the market urgency has meant that these tests do not have a solid clinical validation, which makes them show divergences in sensitivity and specificity in their use [6].

In Ecuador, 102 rapid tests of various brands and methodologies have been authorized by the Health Authorities so far [9].

The present study evaluates a sample of immunofluorescent and immunochromatographic rapid tests to show their agreement in relation to Chemiluminescence.

Methodology

A diagnostic test evaluation assay was performed to establish the performance of five “rapid” tests (4 immunochromatographic and 1 immunofluorescent tests) for IgG and IgM serology for SARS- CoV-2.

The evaluation was performed by comparing to a YHLO_ iFLASH_1800TM Chemiluminescent analyzer, in a medical laboratory with ISO 9001: 2015 certification and with ISO 15189: 2012 accreditation, in the city of Quito. This device was previously verified according to CLSI EP15 A3, with a CV of 3.3% for repeatability and 5.4% for intermediate precision, within manufacturer's recommendations (<10%) [10-12]. Additional performance characteristics are shown in Table 1 [13-14].

Characteristic SARS-CoV-2 IgG SARS-CoV-2 IgM
Detection limit 6 AU/mL 5 AU/mL
Analytical specificity No interference with CMV, M Pneumonia, Chlamydia pn, Rubella, EB VCA, EBNA IgG
Diagnostic sensitivity* 97.30% 86.10%
Diagnostic specificity* 96.30% 92.20%

Table 1: Performance characteristics. YHLO- iFLASH_1800 TM [11-14].

The results obtained from the rapid tests were compared against reports from the chemiluminescent analyzer, using a panel of 30 serum samples from patients received in the laboratory analysis routine. The samples were taken by venepuncture in the anterior side of the elbow, using a multiple-extraction device (Vacuette®), with Vacuette® 21G needle and BD Vacuntainer® tube, with clot activator and gel separator, for serum collection. Samples were mixed by gentle inversion [15]. The tubes were vertically placed on a rack for 30 minutes at room temperature and then were centrifuged at 2000 g for 10 minutes. All samples were free of hemolysis, lipemia, and jaundice.

The signals obtained by the automated analyzer for IgG and IgM were classified as Positive where >10 AU/mL (13/30) for IgG and (6/30) for IgM [11,12].

For the evaluation of clinical performance, the qualitative results of the "rapid" tests were compared against those obtained by chemiluminescence, dichotomized as positives (>10 AU/mL) or negative (<10 UA/mL). A total of 5 "rapid" tests were compared, 4 of them immunochromatographic and 1 fluorescent immunoassay.

As there is no state-of-the-art “reference standard”, this study does not state sensitivity or specificity, nor predictive values, but only the demonstration of agreement in defining the subjects with the dichotomous criterion (positive or negative), by calculating complex repeatability (global percentage of agreements), percentage of positive agreements and of negative agreements, together with their corresponding 95% confidence interval and complemented with Cohen's Kappa test as a measure of agreement that discriminates between agreements by chance [16].

With the obtained data, a Microsoft Excel database was created for subsequent refining and analysis by means of the JASP® software. The luminescent chemo signals detected in the samples were expressed in means and standard deviations, while the concordance were provided in percentages, accompanied by their corresponding confidence interval and through the Cohen's Kappa index [16].

The Research Committee of the Faculty of Medical Science of the Central University of Ecuador review the study and conclude that ethical approval was not required, because the samples of patient are used in the context of clinical diagnoses, and do not exist intervention that affect at clinical decision over patients or their integrity.

Patient and public involvement: Patients and/or the public were not involved in the design, or conduct, or reporting, or dissemination plans of this research.

Results

The characterization in AU/mL for IgG and IgM (AU/mL) for the positive and negative samples is shown in Table 2.

Indicators IgG (AU / mL) IgM (AU / mL)
+ - + -
n 13 17 13 17
Average 51.98 2.899 215.2 1.451
Median 53.07 1.47 8.38 1.16
Standard deviation 19.87 2.924 456.1 1.166
2.5th percentile 10.23 0.33 1.12 0.52
97.5th percentile 86.49 9.03 1407 5.58

Table 2: Characterization (AU/mL) for positive and negative samples IgG and IgM for SARS-COv-2].

Table 3 summarizes the concordance findings of the diagnostic tests evaluated (n=5) in reference to Chemiluminescence.

Maker Type Measurement principle Comparison n * Manufacturer IgG IgM
Diagnostic sensitivity Diagnostic specificity Complex repeatability % (95% CI) Kappa 9Positive agreements % (IC95%)5%] 9Negetive agreements % (IC95%) Complex Repeatability Kappa Positive Agreements % (IC95%) Negative Agreements % (IC95%)
Wondfo Bioteh Co Total Antibodies Fluorescent Immunoassay (PCT) 27 NR NR 92.6 (76.6-97.9) 0.85 100 (Nc) 85.7 (63.4-92.7) 66.7 (47.8-81.4) 0.37 100 (Nc) 57.1 (38.1-74.0)
Vazyme Medical Technology Co. Ltd IgG-IgM Immunochromatography 30 NR NR 80 (62.7-90.5) 0.57 53.8 (88.5-93.0) 100 (Nc) 86.7 (70.3-94.7) 0.63 83.3 (50.5-88.5) 87.5 (70.8-93.8)
Xiamen AmonMed Biotechnology Co., Ltd IgG-IgM Immunochromatography 30 NR * NR * 70 (52.1-83.3) 0.33 30.8 (15.5-91.8) 100 (Nc) 83.3 (66.4-92.7) 0.24 16.7 (9.8-88.5) 100 (Nc)
Shenzhen Rongjin Technology Co. Ltd IgG-IgM Immunochromatography 27 93.30% 96.60% 74.1 (55.3-86.8) 0.49 83.3 (58.9-91.8) 66.7 (44.1-82.4) 66.7 (47.8-81.4) 0.32 83.3 (50.5-88.5) 61.9 (42.4-77.7)
Wondfo Bioteh Co Total antibodies Immunochromatography 30 86.43% 99.57% 80 (62.7-90.5) 0.61 100 (Nc) 64.7 (43.3-80.7) 56.7 (39.2-72.6) 0.25 100 (Nc) 45.8 (29.2-63.7)

Table 3: Performance evaluation, "rapid" serological assays using immunochromatographic and immunofluorescent methods vs. chemiluminescent serology.

Discussion

Given the epidemiological urgency and the growing demand for tests that could contribute to the management of the SARS-CoV pandemic, there has been a high demand for diagnostic tests, including immunoassays, which have had rapid development and commercialization with limited validation in clinical samples [17].

Given the important penetration of rapid tests, mostly immunochromatographic in Ecuador and which have received marketing authorization by the Ministry of Public Health, the present study evaluated the concordance of these tests with a luminescent chemoimmunological analysis, using patients’ samples received at the laboratory for seroprevalence evaluations [16].

The best agreement (both positive and negative agreements) is seen in the immunofluorescent assay, for the IgG contrast, with a particularly good kappa index (0.85), without positive disagreements and a negative disagreement of about 15%. This contrasts with immunochromatographic methods where the Kappa index was 0.61 at best, with disagreements in negative findings of ≈35% and in positive cases of up to ≈70%.

The IgM concordance behavior, on the other hand, reflects a weak to moderate Kappa concordance value (Kappa 0.2 to 0.6), with negative disagreements reaching up to 55% and positives of up to 84%, without any evaluated test reaching Kappa performance equal to or greater than 0.8.

The variations found may be due to the type of antigens used for the development of the assay. Apparently, according to several publications, if they are oriented to nucleocapsid antigens they would seem to be more sensitive, but if they are oriented to the host binding protein (RBD-S) they would be more specific [18].

The great variation in agreement percentages found should draw attention to the problems of diagnostic certainty of serological tests associated with cross-reactions with other coronaviruses. Added to this is the low proportion of negative agreements and their impact on an erroneous screening of a subject for epidemiological surveillance, which may overestimate the population rate considered as immune, when in fact it is not [6].

So far, most of these studies show that people who have recovered from an infection have antibodies to the virus. However, some of these persons have very low levels of neutralizing antibodies in their blood. Therefore, until now, it has not been evaluated whether the presence of detectable antibodies against SARS-CoV-2 confers immunity to a subsequent infection [8]. Serological studies should be used in the clinical and epidemiological context and of other diagnostic tests [8,19].

Conclusion

Given the high demand and supply in the market of "rapid serological tests", its evaluation against panels of serologically positive or negative samples established by Chemiluminescence or Electro chemiluminescence is essential to authorize its extensive use in populations. If use is required, it is recommended that they be carried out by trained technical-operational personnel and under the supervision of professionals in laboratory medicine.

Contributors

Thanks, at Santiago Chicaiza, BF for their help in the selection of the appropriate patient samples required for the implementation of the assay.

References

  1. WHO. Laboratory testing for coronavirus disease 2019 (COVID-19) in suspected human cases. 2020;1-7.
  2. Nguyen T, Duong BD, Wolff A. 2019 Novel Coronavirus Disease (COVID-19): Paving the road for rapid detection and point-of-care diagnostics. Micromachines. 2020;11(3):1-7.
  3. Lippi G, Simundic A-M, Plebani M. Potential preanalytical and analytical vulnerabilities in the laboratory diagnosis of coronavirus disease 2019 (COVID-19). Clin Chem Lab Med. 2020.
  4. Gandhi M, Yokoe DS, Havlir DV. Asymptomatic transmission, the achilles’ heel of current strategies to control COVID-19. New England J Med. 2020;1-3.
  5. World Health Organization (WHO). Coronavirus disease 2019: Situation report -99. 2020;7.
  6. Kontou PI, Braliou GG, Dimou NL. Antibody tests in detecting SARS-CoV-2 infection: A meta-analysis. 2020.
  7. Chen W, Xu Z, Mu J, Yang L, Gan H, Mu F, et al. Antibody response and viraemia during the course of severe acute respiratory syndrome (SARS)-associated coronavirus infection. J Med Microbiol. 2004; 53:435-438.
  8. World Health Organization (WHO). "Immunity passports" in the context of COVID-19: Scientific Brief. 2020; 19-22.
  9. Agencia Nacional de Regulación Cy VS. Pruebas rápidas COVID-19. Pruebas Rápidas Autorizadas COVID-19. 2020.
  10. CLSI C, LSI. CLSI EP15-A3: User verification of precision and estimation of bias. 2014;34: 12.
  11. Shenzen YHLO Biotech co. SARS-CoV-2 IgG iFlash SARS-CoV-2 IgG. 2020; 2-5.
  12. Shenzen YHLO Biotech co. SARS-CoV-2 IgM iFlash SARS-CoV-2 IgM. 2020; 3.
  13. CO SYB. Analytical performance evaluation of four cartridge-type SARS-CoV-2 IgM. 2020; 50.
  14. Shenzen YHLO Biotech co. Analytical performance evaluation of four cartridge-type SARS-CoV-2 IgG. 2020; 50.
  15. Simundic AM, Bölenius K, Cadamuro J, Church S, Cornes MP, van Dongen-Lases EC, et al. Joint EFLM-COLABIOCLI recommendation for venous blood sampling. Ann Biol Clin (Paris). 2019;77(2): 131-154.
  16. CLSI/NCCLS. CLSI EP12-A2. User protocol for evaluation of qualitative test performance; approved guideline. Clin Lab Standards Inst. 2008;48: 64.
  17. Lassaunière R, Frische A, Harboe ZB, Nielsen AC, Fomsgaard A, Krogfelt KA, et al. Evaluation of nine commercial SARS-CoV-2 immunoassays. medRxiv. 2020; 20056325.
  18. To KKW, Tsang OTY, Leung WS, Tam AR, Wu TC, Lung DC, et al. Temporal profiles of viral load in posterior oropharyngeal saliva samples and serum antibody responses during infection by SARS-CoV-2: An observational cohort study. Lancet Infect Dis. 2020;20(5):565-574.
  19. Sharfstein JM, Beckers SJ, Mello MM. Diagnostic testing for the novel coronavirus. J Am Med Assoc. 2020;323(15): 1437-1438.

Author Info

Sáenz-Flor K1,2* and Santafé Lorena M2
 
1School of Medicine, Central University, Quito, Ecuador
2Synlab Solutions in Diagnostics, Quito, Ecuador
 

Citation: Sáenz-Flor K, Lorena MS (2021) Concordance of "Rapid" Serological Tests and IgG and IgM Chemiluminescence for SARS-COV-2. J Antivir Antiretrovir. 13:226.

Received: 04-Jun-2021 Accepted: 18-Jun-2021 Published: 25-Jun-2021 , DOI: 10.35248/1948-5964.21.13.226

Copyright: © 2021 Sáenz-Flor K, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.