Research Article - (2016) Volume 0, Issue 0
A simple, precise and accurate stability indicating RP-HPLC method was developed for the estimation of Alcaftadine in bulk drug and its ophthalmic dosage form. Chromatographic separation of target analyte and degradation products was achieved using Enable HPLC ODS C18 G (250 × 4.6 mm, 5 μm) column and mobile phase comprising of methanol: water, 50:50% v/v at a flow rate of 1.2 ml/min with UV detection at 282 nm. The correlation coefficient (r2) was found to be 0.999 in the concentration range of 1-16 μg/ml. The retention time was found to be 3.15 min. The limit of detection and limit of quantitation was found to be 0.25 μg/ml and 0.75 μg/ml respectively. Stress degradation of Alcaftadine was carried out under stress conditions like acid and base hydrolysis, oxidation, thermal and photolytic stress. The degradation products generated as a result of stress did not show any interference to the detection of Alcaftadine. The developed method was validated, and selectivity of the method was demonstrated by its ability to quantify the target analyte in presence of its degradation products.
Keywords: Alcaftadine; Lastacaft; HPLC; Stability indicating method
The IUPAC name of Alcaftadine is 6,11-dihydro-11-(1- methyl-4-piperidinylidene)-5H-imidazo [2, 1-b][3] benzazepine-3- carboxaldehyde (Figure 1). Alcaftadine is an ophthalmic dual-acting H1-antihistamine and mast cell stabilizer approved for the prevention of itching associated with allergic conjunctivitis. The drug was approved by USFDA in July 2010. It is commercially marketed under the name LASTACAFT. Lastacaft is a sterile, topically administered H1 receptor antagonist containing Alcaftadine for ophthalmic use [1,2].
Alcaftadine is not official in any Pharmacopoeia. A literature survey on Alcaftadine revealed that, until now no analytical method was reported for its determination in bulk drug and its ophthalmic dosage form. However, a clinical pharmacological review report found during the survey showed that liquid chromatography with tandem mass spectrometry (LC/MS/MS) was used to quantitate concentrations of Alcaftadine and R90692 (active metabolite) in K3 EDTA human plasma [3]. It was also found that the metabolic fate of 14 C- Alcaftadine was determined by high performance liquid chromatography-based separation of parent compound from metabolites [4]. Since literature did not cite any method for determination of this drug from bulk drug as well as its formulation it was planned to develop a RP-HPLC method to determine alcftadine in presence of its degradation products.
Instrumentation
Shimadzu Prominence UFLC LC-20AD with UV detector and Enable C18 G RP column (250 × 4.6 mm, 5 μm) was used was chromatographic resolution of target analyte and degradation products. The acquisition and integration of data was performed using LC solution software.
Materials and reagents
Alcaftadine pure drug sample was obtained from JHP Pharmaceuticals LLC, Rochester, Michigan, USA. Lastacaft, Alcaftadine ophthalmic solution, was procured from MediPrime Pharmacy, Dubai. HPLC grade methanol and HPLC water was procured from Merck.
Solution preparation
Standard stock solution: Accurately 10 mg of drug was weighed and transferred to a 10 ml volumetric flask. It was dissolved using methanol and volume was made up to obtain a concentration of 1000 μg/ml. After sufficient dilutions, final stock having concentration equivalent to about 50 μg/ml of drug was prepared.
Working standard solution: The standard stock solution was used to prepare working standard solutions of concentrations 1, 4, 8, 12 and 16 μg/ml. Solution having drug concentration of 8 μg/ml was used as a working standard for stress degradation studies. Both the standard and sample solution of 8 μg/ml were estimated at 282 nm and the chromatograms were recorded (Figure 2).
Sample solution
From the ophthalmic solution (alcaftadine 2.5 mg/ml), one ml was withdrawn and transferred to a 10 ml volumetric flask, dissolved and diluted with methanol to get the concentration of 2500 μg/ml. Further dilutions were made to get the final stock having concentration equivalent to 8 μg/ml.
Optimization of chromatographic conditions
Various trials were made for the selection of chromatographic conditions like mobile phase, its ratio and flow rate. Finally, the one giving the best results were optimized. The chromatographic estimation of Alcaftadine and its separation from degradation products was achieved using Enable C18 G HPLC column using mobile phase Methanol: Water 50:50% v/v at a flow rate of 1.2 ml/min. The UV detection was done at 282 nm. The optimized chromatographic conditions are summarized in Table 1.
Chromatographic parameters | Optimized conditions |
---|---|
Stationary phase | Enable HPLC C18 G 120A Column 250 ×4.6 mm, 5μ |
Mobile phase | Methanol: water, 50:50 (%v/v) |
Flow rate | 1.2 ml/min |
Injection volume | 20μl |
Detection wavelength | 282 nm |
Run time | 10 minutes |
Table 1: Optimized Chromatographic Conditions for RP-HPLC.
Validation of proposed method
The method was validated for the parameters like specificity, linearity, precision, accuracy and robustness as per ICH guidelines [5,6].
System suitability testing
Six replicates of drug concentration of 8 μg/ml were injected and the chromatograms were recorded to check with the system suitability parameters [7].
Forced degradation studies
The forced degradation studies were carried out as per ICH Q1A (R2) for hydrolysis, oxidation and thermal stress conditions and ICH Q1 B for photo stability. The stress conditions employed were 1N HCl for acid hydrolysis, 1 N NaOH for base hydrolysis, 3% H2O2 for oxidative hydrolysis. The drug samples were also subjected to thermal stress conditions as well as subjected to UV light to test photo stability [8-15].
Acid and base hydrolysis: Acid and base induced forced degradation was performed by adding an aliquot of stock solution (1 mg/ml) of Alcaftadine to 1 N HCl and 1N NaOH respectively. These solutions were subjected to stress condition of 50°C for 24 hours. The resulting solutions were sufficiently diluted to obtain the concentration of 8 μg/ml.
Oxidation stress: To study the effects of oxidative conditions, aliquot of stock solution (1 mg/ml) was added to 3% H2O2 solution. It was then kept at 50°C for 24 hours. The resulting solution was sufficiently diluted to obtain the concentration of 8 μg/ml.
Thermal stress: The effect of temperature on Alcaftadine solution was studied by subjecting aliquot of stock solution (1 mg/ml) to hot air oven at 80°C temperature for two hours. The final solution of 8 μg/ml was prepared by adequate dilutions.
Photolytic stress: The aliquot of standard stock solution (1 mg/ ml) was subjected to the UV exposure in UV chamber for 30 minutes. The resulting solution was adequately diluted to obtain the final concentration of 8 μg/ml.
Selection of solvent
Since Alcaftadine was soluble in methanol, it was used as a solvent. For dilution purpose also methanol was only used because on further dilutions with water resulted in splitting of peak.
Optimization of mobile phase
Final mobile phase Methanol: water was selected since trials done using ACN: Water caused peak splitting with different ratios. The ratio 50:50% v/v was optimized because altering the ratio to 60:40 or 70:30 resulted in asymmetrical peak. The flow rate 1.2 ml/min was optimized since at lower flow rates, i.e., at 0.8 ml/min or 1 ml/min, peak tailing of more than 2 was observed.
Optimization of chromatographic conditions
The aim was to develop a method so that it can specifically identify the analyte peak in presence of its degradants. The chromatographic separation of Alcaftadine from its degradants was achieved using Enable C18 G RP HPLC column with mobile phase Methanol: Water (50:50% v/v) at a flow rate of 1.2 ml/min and detection wavelength of 282 nm.
System suitability testing
The results of the system suitability parameters are shown in Table 2. The results were found within the acceptable limits. Hence the system was suitable for the proposed method.
S No | Parameters | Observed results(n=6) | Acceptance criteria | Remarks |
---|---|---|---|---|
1 | Theoretical plates (N) | 3088.474 | >2000 | Method passes the system suitability test |
2 | Tailing factor (T) | 1.36 | T ≤ 1.5 | |
3 | Repeatability (% RSD) | 0.28 | %RSD<2 | |
4 | Capacity factor (k’) | 2.58 | K’>2 |
Table 2: Results for System Suitability Test by RP-HPLC.
Validation
Linearity: The linearity was observed in the concentration range of 1-16 μg/ml for Alcaftadine. The regression line equation was plotted between concentration and peak area. The regression equation y=1639x+50431 was obtained from the linearity data. The correlation coefficient was found to be 0.999.
Specificity: The results of specificity are shown in Figure 3. It showed no interference of the excipients present in drug formulation.
Precision: The results of precision studies are summarized in Table 3. The % RSD was found within the acceptable limit, i.e., <2.
Parameters | Results | |
---|---|---|
Calibration curve range | 1-16μg/ml | |
Regression line equation | Y=1639x+50431 | |
Slope | 1639 | |
Intercept | 50431 | |
Correlation coefficient (r2) | 0.999 | |
Summary of Validation | ||
Accuracy | % Recovery | % RSD |
80% | 98.87 | 0.495 |
100% | 98.95 | 1.340 |
120% | 99.46 | 0.330 |
Precision | Concentration(μg/ml) | % RSD |
Repeatability (n=6) | 8 | 1.205 |
Intraday precision (n=3) | 4 | 0.986 |
8 | 0.920 | |
12 | 1.08 | |
Interday precision (n=3) | 4 | 0.982 |
8 | 0.582 | |
12 | 0.780 | |
LOD | 0.25μg/ml | |
LOQ | 0.75μg/ml | |
Robustness | No significant changes observed with deliberate changes in the method parameters |
Table 3: Summary of the developed and validated HPLC Method.
Recovery studies: The % recovery was observed within the acceptable limits, i.e., 98.71%, 99.06% and 99.39% at the levels of 80%, 100% and 120% respectively. The results are summarized in Table 3.
LOD and LOQ: The results of LOD and LOQ were found to be 0.25 μg/ml and 0.75 μg/ml respectively. The results are summarized in Table 3.
Robustness: Robustness of the method was studied by making variations in the parameters like flow rate (± 0.2 ml/min), mobile phase composition (± 0.2) and detection wavelength (± 2 nm). The deliberate changes made in the flow rate, mobile phase composition and wavelength did not show major impact on retention time, assay value and peak area.
Forced degradation studies
Both the drug (API) as well as formulation was subjected to various stress conditions like acid/ base hydrolysis, oxidation, thermal and photolytic stress conditions as per ICH guidelines. The results are tabulated in Table 4.
Stress conditions | % Degradation | |
---|---|---|
API | Formulation | |
Acid hydrolysis (1N HCl) 50°C, 24 hours | 14.88% | 14.32% |
Base hydrolysis (1N NaOH) 50°C, 24 hours | 15.23% | 14.28% |
Oxidation (3% H2O2) 50°C, 24 hours | 6.89% | 6.35% |
Thermal (80°C, 2 hours) | 15.13% | 16.52% |
Photostability (UV chamber, 30 mins) | 26.32% | 24.02% |
Table 4: Results of Forced Degradation Studies.
Acid and base hydrolysis: The drug degraded in the applied stress conditions of the acid and base hydrolysis i.e., 14.88% and 15.23% respectively for API and 14.32% and 14.28% respectively for formulation. The chromatograms showed the presence of degraded products (Figures 4 and 5).
Oxidation degradation: The drug was found to be quite stable in oxidation stress condition. It showed 6.89% and 6.35% for API and formulation respectively (Figure 6).
Thermal degradation: Adequate thermal degradation was achieved under thermal stress i.e., 15.13% and 16.52% for API and formulation respectively (Figure 7).
Photolytic degradation: The rate of photolytic degradation was higher as compared to others. The drug was degraded sufficiently i.e., 26.32% and 24.02% for API and formulation respectively (Figure 8).
Analysis of the marketed formulation
The proposed method was applied to the marketed formulation of Alcaftadine (LASTACAFT ophthalmic solution, 2.5 mg/ml). The assay results were found to be within acceptable limits, i.e., 99.10-100.39%. The results are shown in Table 5.
Formulation | Labeled amount ml/ml | Amount found mg/ml | % Label claim ± SD |
---|---|---|---|
LASTACAFT(ophthalmic solution) | 2.5 | 2.49 | 99.90 ± 0.698 |
Table 5: Results for the Assay of Ophthalmic Solution of Alcaftadine.
The developed RP-HPLC method for determination of alcaftadine from ophthalmic preparations is specific, reliable and accurate. The observed results also showed the value of relative standard deviations below 2, which indicate the accuracy and precision of the developed method. The proposed method is capable of identifying the target analyte in presence of its degradants. The degradation products generated during stress conditions were not further characterized.
The authors are heartily indebted to Shri D D Patel, President, Om Gayatri Education and Charitable Trust and Pioneer Pharmacy Degree College for their valuable guidance and support. Authors are grateful to the JHP pharmaceuticals, LLC, Rochester, Michigan, USA for providing the drug sample and the Management of Pioneer Pharmacy Degree College for providing necessary research facilities.