In-vivo Corneal Temperature during Cross-linking Measured by an Infrared Thermometer

Hicham Arsalane*, Hamza Elorch, Zakariae Jebbar and Amina Berraho
Department of Ophthalmology, B at Rabat University Hospital, Morocco
*Corresponding Author: Hicham Arsalane, Department of Ophthalmology, B at Rabat University Hospital, Morocco, Tel: +212 661943940 Email:

Keywords: Cornea; Crosslinking; Temperature; Infrared thermometer; Corneal surface; Keratocônus; Riboflavin; UV-A

Introduction

The CXL technique was discovered by Dr. Théo Seller's team at the University of Dresden in the late 1990s while studies in humans began in 2003 [1].The goal of this treatment is to stop progressive and irregular changes in the shape of the cornea known as ectasia. These ectatic changes are typically marked by corneal thinning and increased anterior and/or posterior curvature of the cornea, and often lead to high levels of myopia and astigmatism. The most common form of ectasia is keratoconus [1-5].

This technique requires the addition of a molecule of riboflavin (also called vitamin B2) and the irradiation of corneal tissue with ultraviolet A (UVA) photons. Riboflavin must impregnate the corneal stroma [6-9]. Irradiated by UVA (particularly energetic radiations), this molecule generates free radicals containing oxygen, which would be at the origin of the creation of covalent bonds [1,2,10,11]. Crosslinking of corneal collagen is therefore intended to "stiffen" a biomechanically unstable cornea [10,11]. The principle is based on a photo-induced biochemical "bridging" of collagen fibers [1,2,5,7].

The main goal of the first stage of therapy is to allow riboflavin to diffuse into the cornea. The techniques used to accomplish this all involve either eliminating or weakening the epithelial barrier of the cornea. Conventionally, cross-linking requires central deepithelialization in an operating room with strict sterilization conditions. It is a source of postoperative pain and sometimes of infection [12-16].

The objective of the CXL epi-on is therefore the treatment of keratoconus without pain and without complications related to the deepithelialization: infections, infiltrates, corneal edema, delayed healing, etc. [17-20].The addition of passage-facilitating molecules, such as EDTA or BAC, or a combination of both, breaks the intercellular junctions to allow riboflavin to pass into the stroma [4-7]. This transepithelial technique is the one used in our ophthalmology department B, at RABAT CHU.

Regarding ultraviolet radiation, it is an invisible radiation that emits in the wavelength range of 100 to 400 nm. It has a shorter wavelength than visible light and therefore contains more energy [8,9]. UV radiation is known to be harmful to the endothelium, lens and retina.

Regarding UVA during Cross-Linking sessions [1,2,4,6]:

• The surface irradiance clinically used is 3 mW/cm2.

• The wavelength used is 365 nm with a cumulative illumination of 5.4 J/cm2.

• The duration is 30 min.

Many patients express their concern about the exposure of their cornea to such exposure to UVA radiation and also some ophthalmologists have questions about the safety of their patients. It is true that the appearance of the cornea during the procedure is quite spectacular and fears a burn of the cornea (Figure 1).

clinical-ophthalmology-cross-linking

Figure 1: Aspect of the eye during exposure to UVA radiation in corneal cross-linking.

It is therefore legitimate to ask: what is the temperature of the cornea during this exposure to UVA? And that's the purpose of this study.

Methods

In this prospective study was carried out at the ophthalmology department B at Rabat University Hospital from February 2017 to November 2017. It was proposed to measure corneal temperature using a non-contact infrared thermometer (Benetech gm320)* during crosslinking sessions in patients with evolutive keratoconus. Exclusion criteria were: patients with fever and pachymetry less than 400 μm. This study and data collection complied with all national laws and informed consent was obtained from patients (or their parents for minors) and the study was in accordance with the principles of the Helsinki Declaration.

To know the temperature of an object, the thermometer is pointed towards it and the trigger is pressed. The LCD screen with backlighting allows us to get data even in the dark, as is in case of crosslinking room conditions. It has a laser pointer that was not used in our study. Its measurement range is -50 to +380°C. The distance does not interfere with the temperature. Changing the distance only changes the diameter of the measured surface. The temperature of the room was set at 22°C.

The device was calibrated several times with the mercury thermometer while measuring body temperature, and also with a professional thermometer, the UT151E Modern Digital Multimeters* (figure 2). LIGHTLINK CXL UV-A device was used for this study.

clinical-ophthalmology-infra-red

Figure 2: Calibration of the infra-red thermometer.

A first study concerned 48 normal volunteers (96 eyes), of different age (between 12 And 70 years) and gender (27 females and 21 males), apyretic, all possessing a pachymetry between 520 and 550 μm, in order to have the average corneal temperature which served as a reference for our study. The average temperature between the two eyes was recorded for each volunteer. The measurement of the temperature then concern 46 eyes of 46 patients, apyretic, presenting a evolutive keratoconus with an average pachymetry of 453 μm, benefiting in our service of CXL sessions epi-on with the standard protocol (energy: 3 Mw, duration: 30 min). The average age of patients at the time of the procedure was 19.4 years. The temperature was taken at a fixed distance every 5 min from the time of application of riboflavin on the cornea until 30 min after the end of the CXL session.

A Pearson correlation analysis was done to establish the relationship between two continuous variables, such as corneal temperature and age or pachymetry. The coefficient (r) thus refers to the Pearson correlation coefficient in this article. For all analyzes, the level of significance was set at p<0.05. The analysis was conducted using commercially available software (STATA).

Results

Mean corneal temperature for the whole control group was 34.1 ± 0.6°C (N =48). There was no statistically significant correlation between age and corneal temperature (r=-0.17; N=48; P=0.248017). The result is not significant at p<0.05 (Figure 3).

clinical-ophthalmology-healthy-subjects

Figure 3: Correlation between age in years and corneal temperature in °C in the control group of healthy subjects. There is no statistically significant correlation between age and corneal temperature.

There was no relationship between pachymetry and the temperature of the cornea (N=46; r=-0.09; P=0.551964). The result is not significant at p<0.05 (Figure 4).

clinical-ophthalmology-corneal-temperature

Figure 4: Correlation between pachymetry and corneal temperature in °C in the control group. There is no significant correlation between pachymetry and the temperature of the cornea.

The group of the 48 patients with keratoconus is elaborated with the different parameters: body temperature, age, gender, temperature before the T0 procedure is indicated, then every 5 min during riboflavin drops T1 to T6, and every 5 min also during the irradiation UVA, from T7 to T12. The temperature is taken at 15 min and 30 min after the procedure. The measured pachymetry is also indicated.

The average temperatures of the whole group were measured every 5 min (Figure 5).

clinical-ophthalmology-crosslinking-procedure

Figure 5: Profile changes in corneal temperature during T0 crosslinking procedure (before application of Riboflavin) then every 5 min during instillation of riboflavin (T1 to T6) and every 5 min during UV-A exposure (from T7 to T12). The temperature is taken also, at 30 min after the procedure.

During the riboflavin administration phase, there is little or no increase in temperature. On the other hand, during the UV exposure phase, the temperature of the cornea is increased by an average of 1°C.

So, according to this study (Figure 6):

clinical-ophthalmology-crosslinking-procedure

Figure 6: Evolution of the corneal temperature during the crosslinking session from T0 (before the start of the procedure), at T6 (at the end of the instillation of riboflavin), at T12 (at the end of the UV-A exposure), finally, at 30 min after the radiation. The red line indicates the average temperature of the control group.

• The corneal temperature curves showed a very slight increase when applying Riboflavin alone (+0.2°C).

• But when exposed to UV-A the temperature increased on average by + 1.1°C.

• The temperature returns to the initial values 30 min after the end of UVA irradiation.

Discussion

Studies that also measured infrared temperature of the cornea used an infrared camera [21-25]. It gives color ranges by temperature while the device used in our study targets a well-defined area and displays a very accurate result. Apart from precision, the price difference is also huge (the device used in our study cost only 35 USD, while the entrylevel infrared camera starts at 3000 USD).

The wavelength of UVA used during crosslinking is between 360-370 nm. The energy in joules of the UVA radiation can be calculated from the following relationship: the wavelength in nanometer=celerity of the light (in km/s) divided by the frequency in THerz. The wavelength is 365 nm and the celery of the light is 3,00,000 km per second. So we can find the frequency. To calculate the energy in joule:

E (in joules)=h (Plank constant) x frequency.

With a wavelength of 370 nm and an irradiance of 3 mW/cm2 for a total duration 30 min, this corresponds to a total dose density of 5.4 J/cm2 [26].

1 joule is the energy needed to raise the temperature of one liter of dry air by 1°C.

The wavelength of the UV light used at 370 nm is not chosen at random: it is a wavelength which corresponds to the maximum absorption of riboflavin. Riboflavin (vitamin B2) is not just a photosensitizer, it also acts as a UV absorber. Because of the extra shielding of riboflavin, all the structures located behind the corneal stroma, including the corneal endothelium, the anterior chamber, the iris, lens and retina are theoretically exposed to a residual density less than 1 J/cm2 [20].Moreover, no retinal or crystalline involvement after CXL has been described in the literature [20].

This study reinforces the veracity of the safety of the procedure, because the increase in the temperature of the cornea after exposure to UVA crosslinking for 30 min, remains about 1°C. This increase in temperature is identical to that which occurs physiologically when closing the eyelids for 5 min [27,28].

Conclusion

Although this study is limited by the small number of patients studied and is performed only on cases of crosslinking epi-on, it is clear that there is an increase in corneal temperature of about 1°C during the crosslinking sessions with the standard strategy.

So contrary to popular belief about exposure of the cornea to irradiation by UV-A, not only by patients, but also by some ophthalmologists, there is increase in corneal temperature which remains limited to 1 degree Celsius.

In anticipation of a broader study with different crosslinking strategies, this study has the merit of introducing this very cheap, noninvasive and accurate tool for measuring corneal temperature in-vivo .

References

  1. Wollensak G, Spoerl E, Seiler T (2003) Riboflavin/ultraviolet-a–induced collagen crosslinking for the treatment of keratoconus . Am J Ophthalmol 135: 620-627.
  2. Wollensak G (2006) Crosslinking treatment of progressive keratoconus: new hope. Curr Opin Ophthalmol 17: 356-360.
  3. Gore DM, Shortt AJ, Allan BD (2013) New clinical pathways for keratoconus. Eye (Lond) 27: 329-339.
  4. Pron G, Ieraci L, Kaulback K (2011) Collagen cross-linking using riboflavin and ultraviolet-A for corneal thinning disorders : an evidence-based analysis. Ont Health Technol Assess Ser [Internet] 11: 1-89.
  5. Erdem Y, Bektas C, Bilgihan K (2015) Transepithelial Versus Epithelium-off Corneal Cross-Linking for the Treatment of Progressive Keratoconus: A Randomized Controlled Trial. Am J Ophthalmol 160: 399-400.
  6. Hammer A, Tabibian D, Richoz O, Hafezi F (2014) Prise en charge du kératocône par cross-linking du collagène cornéen. Rev Med Suisse 10: 1263-1265.
  7. Hashemi H, Seyedian MA, Miraftab M, Fotouhi A, Asgari S (2013) Corneal collagen cross-linking with riboflavin and ultraviolet A irradiation for keratoconus : Long-term results. Ophthalmology 120: 10515-1520
  8. Zhen-Yong Z (2014) Corneal Collagen Cross-linking With Riboflavin and Ultraviolet-A Irradiation in Patients With Thin Corneas. Am J Ophthalmol 153: 1002.
  9. Meek KM, Tuft SJ, Huang Y, Gill PS, Hayes S, et al. (2005) Changes in collagen orientation and distribution in keratoconus corneas Invest . Ophthalmol Vis Sci 46: 1948-1956.
  10. Raiskup F, Spoerl E (2011) Corneal Cross-linking with Hypo-osmolar Riboflavin Solution in Thin Keratoconic Corneas. Am J Ophthalmol 152: 28-32.
  11. Chan BP, So KF (2005) Photochemical crosslinking improves the physicochemical properties of collagen .scaffolds J Biomed Mater Res A 75 : 689-701.
  12. Kamaev P, Friedman MD, Sherr E, Muller D (2012) Photochemical kinetics of corneal cross-linking with riboflavin. Invest Ophthalmol Vis Sci 53: 2360-2367.
  13. Leccisotti A, Islam T (2010) Réticulation du collagène cornéen transépithélial dans le kératocône. J Refract Surg 26: 942-948.
  14. Dhawan S, Rao K, Natrajan S (2011) Complications of Corneal Collagen Cross-Linking. J Ophthalmol p. 869015.
  15. Pollhammer M., Cursiefen C (2009) Bacterial keratitis early after corneal crosslinking with riboflavin and ultraviolet-A. J Cataract Refract Surg 35: 588-589.
  16. Rama P, Di Matteo F, Matuska S, Paganoni G, Spinelli A (2009) Acanthamoeba keratitis with perforation after corneal crosslinking and bandage contact lens use. J Cataract Refract Surg 35: 788-791.
  17. Zamora KV, Males JJ (2009) Polymicrobial keratitis after a collagen cross-linking procedure with postoperative use of a contact lens: a case report. Cornea 28: 474-476.
  18. Kymionis GD, Portaliou DM, Bouzoukis DI, Suh LH, Pallikaris AI et al. (2007) Herpetic keratitis with iritis after corneal crosslinking with riboflavin and ultraviolet A for keratoconus . J Cataract Refract Surg 33: 1982-1984.
  19. Rubinfeld RS, Rabinowitz YS (2012) CXL With the Epithelium on or off: Which Is Better? Cataract Refract Surg 2012: 1-5.
  20. Spoerl E, Mrochen M, Sliney D, Trokel S, Seiler T (2007) Safety of UVA-riboflavin cross-linking of the cornea. Cornea 26: 385-389.
  21. H Fujishima, I Toda, M Yamada, N Sato, K Tsubota (1996) Corneal temperature in patients with dry eye evaluated by infrared radiation thermometry. Br J Ophthalmol 80: 29-32.
  22. Sniegowski M, Erlanger M, Velez-Montoya R, Olson JL (2015) Difference in ocular surface temperature by infrared thermography in phakic and pseudophakic patients. Clin Ophthalmol 9: 461-466.
  23. Mapstone R (1968) Measurement of corneal temperature. Exp Eye Res 7: 237-243.
  24. Carracedo G, Rodríguez-Pomar C, Martín-Hermoso A, Martin-Gil A, Pintor J (2016) Ocular Surface Temperature and Tear Film Matrix Metalloproteinase-9 Concentration in Sjögren Syndrome Patients. J Clin Exp Ophthalmol 7: 3.
  25. Kessel L, Johnson L, Arvidsson H, Larsen M (2010) The Relationship between Body and Ambient Temperature and Corneal TemperatureInvest Ophthalmol Vis Sci 51: 6593-6597.
  26. Tkáčová M, Živčák J, Foffová P (2011) Reference for Human Eye Surface Temperature Measurements in Diagnostic Process of Ophthalmologic Diseases. Measurement 1:1
  27. Spoerl E, Hoyer A, Pillunat LE, Raiskup F (2011) réticulation cornéenne et les questions de sécurité. Ophthalmol J 5: 14-16.
  28. Mapstone R (1968) Determinants of corneal temperature. Bri J Ophthalmol 52: 729-741.
Citation: Arsalane H, Elorch H, Jebbar Z, Berraho B (2018) in-vivo Corneal Temperature during Cross-linking Measured by an Infrared Thermometer. J Clin Exp Opthamol 9: 758.

Copyright: © 2018 Arsalane H, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.