Emergency Medicine: Open Access

Emergency Medicine: Open Access
Open Access

ISSN: 2165-7548

+44 1223 790975

Case Report - (2015) Volume 5, Issue 5

Inappropriate Stent Placement Resulting in a Coronary Slow-Flow: Case Report

Igor Kranjec1*, Blaz Mrevlje2 and Miha Cercek1
1Department of Cardiology, University Medical Centre, Ljubljana, Slovenia
2Department of Cardiology, University Medical Centre, Rostock, Germany
*Corresponding Author: Igor Kranjec, Jakceva 43, 1000 Ljubljana, Slovenia, Tel: +38641683230 Email:

Abstract

A case of primary percutaneous coronary intervention (PCI) is shown in the elderly patients with an ST-segment myocardial infarction. Stent deployment following the extensive thrombus aspiration resulted in a slow-flow phenomenon. Optical coherence tomography imaging showed a huge thin-capped fibro-atheroma encroaching the landing zone for stent placement. The author discusses possible factors involved in the development of the slowflow.

Keywords: Primary PCI; OCT intravascular imaging; slow-flow; Thin-capped fibro-atheroma; Inappropriate landing zone

Introduction

Primary percutaneous coronary intervention (PCI) is the standard care for patients with an ST-segment elevation myocardial infarction (STEMI) [1]. However, there are some patients who seem not to benefit fully from prompt restoration of antegrade flow, as they fail to show resolution of signs of ischemia and improvement in perfusion abnormalities. These patients present an angiographic phenomenon characterized by evidence of slow-flow (SF) in the affected vessel (Thrombolysis in Myocardial Infarction [TIMI] flow grade =2) and lack of contrast uptake “blush” despite the absence of mechanical obstructions. The prevalence of SF has been reported as 0.2-30% in various studies [2]. SF has a multifactorial pathogenesis including distal embolization, ischemia-reperfusion injury, and individual predisposition. It is possible that the importance of each pathogenetic component is different in various patients, thus explaining the occurrence of SF despite the use of mechanical thrombus aspiration. A series of consistent data has clearly shown that SF has a strong negative impact on outcome, negating the potential benefit of primary PCI [3]. We report, therefore, a case of a patient with STEMI who developed SF despite the extensive thrombus aspiration. Using the optical coherence tomography (OCT), we proposed that SF might have developed as a consequence of inappropriate stent positioning.

Case Report

A 78-year-old lady with hypertension presented to our hospital with severe, persistent chest pain. On admission, her respiratory rate was 12/min, oxygen saturation 93%, heart rate 70/min, blood pressure 125/80 mm Hg, with basal crackles. In the electrocardiogram (ECG), there were marked ST-segment elevations in the leads II, III, aVF, and V6. She was immediately taken to the catheterization laboratory. The emergent coronary angiography revealed a tight (~95%), long lesion with a large filling defect in the middle right coronary artery (RCA); yet distal blood flow was rather preserved (Figure 1a). Left coronary system was unremarkable. At that point, OCT imaging was performed using the C7 Dragonfly catheter and the IlumienTM system (St. Jude Medical Europe, Inc., Zaventem, Belgium).

emergency-medicine-angiogram-right-coronary

Figure 1: a) Angiogram of the right coronary artery (RCA) initially. Note tight stenosis (black arrow) with a haziness representing intraluminal thrombus, b) Angiogram of the RCA after stenting. Note the slow-flow phenomenon (black arrow), c) Shown is the placement of the second stent in the upper RCA. Note the shadow of the already deployed distal stent (black arrow).

The imaging discovered a massive fibro-lipidic plaque with a large lipid pool and a thin fibrous cap (TCFA), extensive intimal dissection, and mixed thrombus with a long tail (Figure 2a). Furthermore, there was another, huge TCFA ~15 mm proximal to the culprit lesion (Figure 2b). Consequently, the primary PCI was carried out under protection of the unfractionated heparin. Initially, the manual aspiration by means of the ExportR 6F catheter (Medtronic Vascular, Santa Rosa, Ca, USA) yielded large pieces of red thrombus. Two bare-metal stents Multi-link Vision (Abbott Vascular, Santa Clara, CA, USA), 3.0×18 mm distally and 3.0×15 mm proximally, were eventually implanted at 18 Atm and stent overlap was post-dilated at 20 Atm. Repeat angiography showed nicely deployed stents though the peripheral blood blood had slowed down to TIMI grade 2 (Figure 1b) immediately after the implantation of the proximal stent (Figure 1c). Repeat OCT imaging revealed an appropriate expansion of both stents (Figure 3a); however, a few struts of the proximal stent were wedged into the detached TCFA that remained partly uncovered (Figures 3 b,c). As the patient was doing well and the ECG hadn’t changed, the PCI was brought to an end without further adjustment. The cardiac troponin I raised to a peak of 20.662 ng/ml (upper limit of normal 0.01 ng/ml) on the same day. The further hospital course was unremarkable and the patient was transferred to the peripheral hospital on the third day, receiving aspirin, clopidogrel, ß-blocker, ACE inhibitor, and statin.

emergency-medicine-oct-image-culprit-lesion

Figure 2: a) OCT image of the culprit lesion beyond the tightest narrowing. Note the intimal dissection (white arrow); thrombus (T) is mostly intraluminal, b) OCT image of the right coronary artery proximal to the culprit lesion. Shown is the thin-capped fibro-atheroma (white arrow); the thinnest part of the fibro cap I seen next to the OCT catheter.

emergency-medicine-oct-image-right-coronary

Figure 3: a) OCT image of the right coronary artery after stenting. Note well expanded and apposed stent struts, b) OCT image of the proximal stent edge. Note the wedged stent struts (white arrow) into the thin-capped fibro-atheroma; there are some malapposed struts at 9 o’clock, c) OCT image of the vessel segment adjacent to the proximal stent edge. Note the thin-capped fibro-atheroma at 7 o’clock. Stent struts are no more visible.

Discussion

We report the case of the elderly patient who had suffered from STEMI caused by the thrombotic obstruction of the RCA. Despite the extensive manual thrombus aspiration, the primary PCI resulted in the SF decreasing the benefit of the achieved reperfusion. The OCT imaging revealed the struts of the proximal stent to be wedged into the detached TCFA.

There has been a growing wealth of scientific literature pointing at the worrying influence of the fibrolipidic plaques, particularly TCFAs, on PCI outcomes. The presence of TCFA by itself has been associated with SF, microvascular obstruction, and peri-procedural myocardial infarction in patients who had undergone elective or urgent PCI [4,5]. Moreover, the volume of the fibro-fatty plaque over the entire length of stenosis, determined by virtual histology intravascular ultrasound imaging, was an independent predictor of SF during primary PCI (beta=0.359, p=0.006) [6]. Tanaka et al. [7] proved that TCFAs, demonstrated by OCT, were more frequently observed in non-STEMI patients who developed PCI-related SF (50% vs. 16% in no-SF patients, p=0.005); furthermore, a multivariable logistic regression analysis revealed that lipid arch wider than 90o was an independent predictor of SF (odds ratio 1.018, p=0.01).

It has been widely known that interventional cardiologists prefer to position stents’ edges into normal or at least “less diseases” vessel references. One of the advantages of OCT guidance is that it can help precisely identify the optimal segment for stent placement – the so-called stent landing zone [8]. Recently reported clinical cases have proven pre-procedural OCT’s unique ability to avoid late stent failure, revealing at the landing zone the presence of inflamed, lipid-rich core plaques (LCP), undetected by angiography and causing local restenosis [9]. TCFAs at proximal stent edges were also predictors of uncovered struts [10]. Finally, Hanson et al. [11] demonstrated using near-infrared spectroscopy that LCPs extended beyond angiographic target lesions in nearly 20% of cases. Disturbingly, the failure to cover the longitudinal extent of lesions, particularly if the uncovered lesion segment contained LCP, was associated with increased rate of stent thrombosis and restenosis.

Large thrombus burden has been demonstrated as an independent predictor of major adverse events and stent thrombosis in culprit arteries of patients treated for STEMI [12]. Distal embolization has been repeatedly shown during primary PCI, particularly after stent deployment [13]. Routine manual thrombectomy, as compared with PCI alone, was not able to reduce the composite of clinical adverse events (6.9% vs. 7.0%, p=0.86) in the recent TOTAL trial [14]. Moreover, manual aspiration didn’t increase OCT-determined flow area in the TROFI study [15]. One of possible explanations for the aspiration failure might be the inefficiency of the current manual device as a5ppreciated in the study of Imola and co-workers [16,17].

Taken together, the scientific data and the presented case altogether point at the deleterious influence of the inappropriate landing zone for stenting on the development of SF in primary PCI. Certainly, contribution of distal embolization must be taken into account. We argue, however, that SF in our case appeared after placing the second, proximal stent which was possibly trying to avoid the tightest part of the lesion (Figure 1c). The largest amount of the thrombus was seen beyond the stenosis as a long tail (Figure 2a). Co-registration of OCT pullback with X-ray angiogram in our case would definitely prove useful to explain in more detail the cause of SF.

References

  1. Authors/Task Force members, Windecker S, Kolh P, Alfonso F, Collet JP, et al. (2014) 2014 ESC/EACTS Guidelines on myocardial revascularization: The Task Force on Myocardial Revascularization of the European Society of Cardiology (ESC) and the European Association for Cardio-Thoracic Surgery (EACTS)Developed with the special contribution of the European Association of Percutaneous Cardiovascular Interventions (EAPCI).Eur Heart J 35: 2541-2619.
  2. Sani SHD, Eshraghi A, Shahri B, Vejdanparast M (2014) No-Reflow Phenomenon in Patients with ST-Elevation Acute Myocardial Infarction, Treated with Primary Percutaneous Coronary Intervention: A Study of Predictive Factors. J Cardiothorac Med 2: 221-226.
  3. Niccoli G, Burzotta F, Galiuto L, Crea F (2009) Myocardial no-reflow in humans.J Am CollCardiol 54: 281-292.
  4. Ozaki Y, Tanaka A, Tanimoto T, Kitabata H, Kashiwagi M, et al. (2011) Thin-cap fibroatheroma as high-risk plaque for microvascular obstruction in patients with acute coronary syndrome.CircCardiovasc Imaging 4: 620-627.
  5. Porto I, Di Vito L, Burzotta F, Niccoli G, Trani C, et al. (2012) Predictors of periprocedural (type IVa) myocardial infarction, as assessed by frequency-domain optical coherence tomography.CircCardiovascInterv 5: 89-96, S1-6.
  6. Bae JH, Kwon TG, Hyun DW, Rihal CS, Lerman A (2008) Predictors of slow flow during primary percutaneous coronary intervention: an intravascular ultrasound-virtual histology study.Heart 94: 1559-1564.
  7. Tanaka A, Imanishi T, Kitabata H, Kubo T, Takarada S, et al. (2009) Lipid-rich plaque and myocardial perfusion after successful stenting in patients with non-ST-segment elevation acute coronary syndrome: an optical coherence tomography study.Eur Heart J 30: 1348-1355.
  8. Karanasos A, Ligthart J, Witberg K, van Soest G, Bruining N, et al. (2012) Optical Coherence Tomography: Potential Clinical Applications.CurrCardiovasc Imaging Rep 5: 206-220.
  9. Ikenaga H, Ishihara M, Inoue I, Kawagoe T, Shimatani Y, et al. (2013) Longitudinal extent of lipid pool assessed by optical coherence tomography predicts microvascular no-reflow after primary percutaneous coronary intervention for ST-segment elevation myocardial infarction.J Cardiol 62: 71-76.
  10. Ueda T, Uemura S, Dote Y, Goryo Y, Sugawara Y, et al. (2014) Placement of Proximal Edge on TCFA and Calcification Correlates With Inadequate Neointimal Coverage of Struts After Drug Eluting Stent Implantation. Circulation 130: A12673.
  11. Hanson I, Dixon S, Goldstein J (2013) Employing NIRS-IVUS to Guide Optimal Lesion Coverage—Avoidance of Geographic Miss. J Invasive Cardiol 25:18A-19A.
  12. Sianos G, Papafaklis MI, Daemen J, Vaina S, van Mieghem CA, et al. (2008) Angiographic stent thrombosis after routine use of drug-eluting stents in ST-segment elevation myocardial infarction: the importance of thrombus burden. J Am CollCardiol 50:573-83.
  13. Okamura A, Ito H, Iwakura K, Kurotobi T, Koyama Y, et al. (2008) Clinical implications of distal embolization during coronary interventional procedures in patients with acute myocardial infarction: quantitative study with Doppler guidewire.JACC CardiovascInterv 1: 268-276.
  14. Jolly SS, Cairns JA, Yusuf S, Meeks B, Pogue J, et al. (2015) Randomized trial of primary PCI with or without routine manual thrombectomy.N Engl J Med 372: 1389-1398.
  15. Onuma Y, Thuesen L, van Geuns RJ, van der Ent M, Desch S, et al. (2013) Randomized study to assess the effect of thrombus aspiration on flow area in patients with ST-elevation myocardial infarction: an optical frequency domain imaging study--TROFI trial.Eur Heart J 34: 1050-1060.
  16. Imola F, Mallus MT, Ramazzotti V, Manzoli A, Di Vito L, et al. (2015) Quantification of manual thrombus removal in patients with acute coronary syndromes: a study exploiting serial frequency domain-optical coherence tomography. J Cardiovasc Med (Hagerstown) 16: 204-212.
  17. Hebsgaard L, Nielsen TM, Tu S, Krusell LR, Maeng M, et al. (2015) Co-registration of optical coherence tomography and X-ray angiography in percutaneous coronary intervention. The Does Optical Coherence Tomography Optimize Reva scularization (DOCTOR) fusion study. In J Cardiol 182: 272-278.
Citation: Kranjec I, Mrevlje B, Cercek M (2015) Inappropriate Stent Placement Resulting in a Coronary Slow-Flow: Case Report. Emerg Med (Los Angel) 5:275.

Copyright: © 2015 Kranjec I, et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Top