Objective: To investigate the impact of a reconstruction method incorporating corrections for scatter, attenuation, and distance dependent detector response on cardiac single emission computed tomography (SPECT) perfusion studies compared with filtered back projection (FBP).
Materials and methods: A total of 20 patients underwent same-day, rest-stress SPECT/CT myocardial perfusion imaging. Images were reconstructed using iterative 3D ordered-subsets expectation-maximization (OSEM 3D) and FBP algorithms. Stress and rest myocardial perfusion defects were quantified using polar maps and normal database comparisons. The Bland-Altman plots were used to assess their degree of agreement. Results were confirmed by coronary angiography. The contrast, contrast to noise ratio and signal to noise ratio were used for the quantitative evaluation of the reconstruction quality.
Results: Perfusion defect extent quantification on OSEM 3D reconstructed images agreed and correlated well with defect extent quantification on FBP reconstructed images (bias ± standard deviation, -15% ± 20; r = 0.63) during stress and at rest (-10% ±15; r=0.70). Agreement and correlation were similar for severity scores during stress (-1.02 ± 1.77 SD’s; r=0.62), and at rest (-1.10 ± 1.49 SD’s; r=0.61). There were no statistically significant differences between the methods regarding perfusion defect extent or severity. The overall agreement rate with coronary angiography was similar. OSEM 3D reconstruction algorithm significantly increases the image contrast by 31% (P<0.05).
Conclusions: Compensation for detector response, attenuation and scatter improves image contrast compared with FBP. Applying quantitative analysis, OSEM 3D reconstruction produced increased image contrast compared to FBP, but similar results regarding the size and severity of left ventricular perfusion defects.
Keywords: Myocardial perfusion; Cardiac SPECT; OSEM 3D reconstruction; FBP reconstruction; Coronary angiography
Myocardial perfusion imaging (MPI) using single-photon emission tomography (SPECT) is a widely used noninvasive modality in the diagnosis of coronary artery disease (CAD) and for assessing the functional significance of coronary stenoses [1,2]. An adequate image reconstruction algorithm is essential for image quality and diagnostic value of the technique. The routine for cardiac SPECT image reconstruction that has been used for many years was filtered backprojection (FBP) [3]. It is computationally fast, easy to implement, and widely clinically applied. Nevertheless, the FBP reconstructed images are degraded by a number of factors including soft tissue attenuation, Compton scatter, and collimator blurring, which limit spatial resolution, degrade the image contrast, and may distort the shape of left ventricular (LV) myocardium [4]. The introduction of hybrid SPECT/CT scanners in nuclear cardiology offers the potential of performing CT-based attenuation correction (AC) of the emissions scans, thus optimizing the diagnostic capabilities of cardiac SPECT. New reconstruction algorithms have been developed to include attenuation correction in the reconstruction process, as well as other parameters, such as depth-dependent collimator blurring and scatter [5-7]. Recent work has indicated that iterative reconstruction (IR) algorithms can be used to model the spatial resolution and sensitivity and compensate for the effects of scatter and attenuation when using CT [8-10]. Currently, ordered-subset expectation maximization (OSEM) is the preferred iterative method for reconstructing cardiac SPECT studies. Published data have demonstrated the ability of OSEM to improve lesion detectability, overall image quality and contrast, indicating a substantial advantage over the conventional FBP method [11,12]. This is particularly important for the interpretation of image data affected by visceral activity, breast tissue attenuation in women, or diaphragmatic attenuation, which can be confused with true perfusion defects.
The current study seeks to evaluate the overall cardiac image quality of the OSEM with three dimensional resolution recovery (OSEM 3D) and CT based attenuation correction as compared to the conventional FBP reconstruction and to correlate the MPI scan findings with the coronary angiography.
Study population
The characteristics of the study population are shown in Table 1. The study involved 20 patients (13 males, 8 females, median age 61 years, range 41–80 years) who underwent MPI followed by invasive coronary angiography (CAG) within 3 months, for any positive or equivocal findings for CAD at MPI (using OSEM 3D reconstruction algorithm). All patients underwent one-day, low-dose rest, high-dose stress Tc- 99m Tetrofosmin myocardial perfusion SPECT. SPECT images were acquired between 15 and 30 minutes after rest injection and between 45 and 60 min after stress injection. All subjects were given 0.14 mg/ kg/min dipyridamole intravenously over 4 min. Three minutes after the end of dipyridamole infusion, the radiopharmaceutical was injected through an antecubital vein. Hemodynamic data and 12-lead ECGs were recorded every minute during dipyridamole infusion. Informed consent was obtained from all patients after they received a detailed explanation of the procedure.
Patient characteristic | Number of patients (%) |
Abnormal rest ECG results | 13 (65%) |
Prior myocardial infarction | 7 (35%) |
Prior catheterization | 7 (35%) |
Revascularization PTCA CABG |
5 (25%) 2 (10%) |
Pacemaker | 2 (10%) |
Hypertension | 9 (45%) |
Diabetes | 2 (10%) |
Table 1: Characteristics of the study population.
Spect/ct imaging protocol
All studies were completed with a 90°-angled dual-head integrated Symbia T6 SPECT/ CT camera equipped with low-energy, highresolution collimators. Patients were placed in the supine position with arms up, and an end expiration CT scan was acquired before SPECT images using a tube current of 130 kV, effective mAs of 13, delay 3s, for an axial field of view covering the thorax from the lung apex to 1-2 cm below the diaphragm. Emission data were acquired with a matrix of 64x64 pixels and a 20% energy window centered at 140 keV. Further acquisition parameters were 32 projections, 180° orbit, and 25 s per projection. On a dedicated workstation (Syngo MMWP, Siemens Medical Solutions) images were reconstructed iteratively (10 iterations, 8 subsets, Gaussian filter, FWHM= 9.6) by using the OSEM 3D approach (e.soft “Flash3D,” Siemens Medical Solutions) with compensation for collimator-detector response as described previously elsewhere [12,13]. Datasets were visually assessed for misregistration with a fusion/registration software tool (Syngo Advanced Fusion, Siemens Medical Solutions), and when needed, a manual realignment between SPECT and CT images was performed to improve the CT attenuation correction process (Figure 1). The raw data were again reconstructed using the FBP method (Butterworth filter, Order 5, with a cut off frequency of 0.45 cycles/cm) at the same reorientation and reslicing angles as was used by Flash 3D. The numbers of iterations and subsets as well as the post smoothing parameters used were those recommended by the software supplier for the acquisition type. (Siemens/4D-MSPECT, INVIA, Ann Arbor, MI).
Figure 1: The resulting image from FBP and Flash 3D reconstruction. Top row shows the alignment of cardiac data matching CT and SPECT images. Middle row shows midventricular short-axis slices reconstructed with FBP. Note the presence of an inferior wall perfusion defect on stress-rest images indicating an inferior wall myocardial infarction. Bottom row shows midventricular short axis slices reconstructed with Flash 3D. Note the absence of the inferior wall defect, suggesting that this was an artifact. Also, better contrast between the left ventricular walls and cavity and better visualization of the right ventricle.
Data analyses
Polar map plots were generated for both stress and rest cardiac tomograms. Normal databases matching the criteria of the selected dataset (gender specific, image acquisition performed in the same physiologic state, specific image based reconstruction technique) were used to define normal myocardial perfusion. Myocardial perfusion defects were automatically identified as areas with counts below the lower limit of normal (2.5 SD’s as compared with the corresponding database). The extent of the perfusion defect was expressed as a percentage of the left ventricular surface area and the severity, as units of standard deviations (SD) below the normal mean. In addition, the correlation between the anatomic localization of the stenotic lesions and the location of detected perfusion defects were assessed using the polar map displays and a coronary artery distribution model template included in the software package (Siemens/4D-MSPECT, INVIA, Ann Arbor, MI).
The image quality of FBP and Flash 3D reconstruction was characterized using three quantitative criteria: contrast (C), contrast to noise ratio (CNR) and signal to noise ratio (SNR). Contrast (C) was calculated between two regions of interest (ROIs) drawn inside the LV wall and cavity. The mean numbers of counts in these two ROIs were calculated, and the contrast was defined as:
,where AMyo and ACavity
are the mean number of counts of the LV wall and the LV cavity, respectively
Noise (σMyo) was estimated as the standard deviation of the mean number of counts of the myocardial ROI.
Contrast to noise ratio (CNR) was determined as the ratio between contrast over the noise:
CNR=C/σMyo
SNR in the myocardium region was calculated as the ratio between the mean number of counts over the noise:
SNR=(AMyo)/σMyo
Statistical analyses
Data from the 2 reconstruction methods were compared using the paired t test. Independent groups were compared using an independent t test or 1-way ANOVA with Bonferonni correction. The Fisher exact test was used for comparison of proportional data. The estimates of left ventricle perfusion defects (percentage of left ventricular defect and severity scores) between the two reconstruction methods were compared using the Bland-Altman method of comparison [14] and Pearson correlation coefficient. A P value of less than 0.05 was considered statistically significant. All statistical tests were performed with MedCalc for Windows (MedCalc Software, Mariakerke, Belgium).
Coronary angiography
The coronary arteries have been found to be normal on invasive coronoary angiography in five of the 20 patients (25%). A significant stenosis (arterial narrowing exceeding 50% of the lumen diameter) was present in the left anterior descending coronary artery (LAD) in 9 cases, left circumflex artery (LCx) in 6 cases and right coronary artery (RCA) in 11 cases. Single vessel disease was present in six patients (30%), double vessel disease in 7 (35%) and triple vessel in 2 (10%) patients.
Characterization of myocardial perfusion defects
The corresponding polar maps of myocardial perfusion were assigned to one of the three major coronary arteries using a standard template. All myocardial perfusion defects were automatically quantified for the extent and severity with both SPECT reconstruction methods (Figure 2). Descriptive statistics for the perfusion defect measurements are shown in Table 2 and Figure 3.
Figure 2: The coronary artery territory template used for defect extent and severity scores quantification with both reconstruction methods.
Variable | Abn. segm (N) | Mean | SD | SEM | 95% CI of Mean | |
Extent % Stress Severity Score | Flash 3D LAD LCx RCA FBP LAD LCx RCA |
17 17 17 12 13 9 |
31.94 29.52 41.88 25.08 25.61 25.00 |
23.58 23.65 27.70 22.44 23.90 23.74 |
5.71 5.73 6.71 6.47 6.62 7.91 |
19.81 to 44.06 17.36 to 41.68 27.63 to 56.12 10.82 to 39.34 11.17 to 40.06 6.74 to 43.25 |
Flash 3D LAD LCx RCA FBP LAD LCx RCA |
17 17 17 12 13 9 |
3.77 3.80 3.75 3.49 3.73 3.52 |
0.89 1.11 0.56 1.19 1.36 0.92 |
0.24 0.29 0.15 0.31 0.39 0.30 |
3.23 to 4.31 3.15 to 4.44 3.43 to 4.08 2.80 to 4.18 2.86 to 4.60 2.80 to 4.23 |
|
Extent % Rest Severity score | Flash 3D LAD LCx RCA FBP LAD LCx RCA |
19 18 19 11 14 11 |
27.47 22.77 26.84 26.54 19.21 24.09 |
21.78 17.57 21.91 22.29 18.71 19.36 |
4.99 4.14 5.02 6.72 5.02 5.83 |
16.97 to 37.97 14.03 to 31.51 16.27 to 37.40 11.56 to 41.52 8.40 to 30.02 11.07 to 37.10 |
Flash 3D LAD LCx RCA FBP LAD LCx RCA |
19 18 19 11 14 11 |
3.39 3.33 3.75 3.56 3.46 3.52 |
1.29 1.34 0.56 1.11 1.08 0.92 |
0.33 0.34 0.15 0.32 0.29 0.30 |
2.67 to 4.10 2.59 to 4.07 3.43 to 4.08 2.85 to 4.27 2.83 to 4.09 2.80 to 4.23 |
Table 2: Comparative description of FBP and Flash 3D derived defect extent (number of segments and in %) and severity.
Perfusion defects on stress images: A total of 60 myocardial segments were included in stress analysis: 51 (85%) showed decreased perfusion with Flash 3D and 34 (57%) with FBP reconstruction. The mean differences in defect extent between the two methods of reconstruction were 6% for the LAD artery, 4% for the LCx and 17% for the RCA artery. Correspondingly, the mean differences in severity scores were 0.3SD’s for the LAD, 0.1SD’s for the LCx and 0.2 SD’s for the RCA. No significant differences were found between the two methods of reconstruction regarding the defect extent and severity during stress. When all defect extent and severity scores of all coronary arteries were pooled, the Bland-Altman method of comparison revealed an observed bias of -15 ± 20.73% (overall correlation coefficient, r=0.63) for the defect extent measurement, and a bias of -1.02±1.77 SD’s (overall correlation coefficient, r=0.62) for the severity score.
Perfusion defects on rest images: Of 60 myocardial segments studied, decreased myocardial perfusion was found in 56 (93%) in Flash 3D based reconstructions and in 36 (60%) in FBP. The mean differences in defect extent between the reconstruction methods were 1%, 4% and 3% within the vascular territory of LAD, LCx, and RCA, respectively. Mean severity scores slightly increase in FBP by 0.2 SD’s within the LAD and by 0.1 SD’s in LCx and decrease by 0.2 SD’s within the RCA vascular territories. No significant differences were encountered between the two methods of reconstruction at rest. Bland-Altman analysis for pooled data showed a bias of -10.27±15.37% (overall correlation coefficient, r=0.70) for the defect extent size, and a bias of -1.10±1.49 SD’s (overall correlation coefficient, r=0.61) for severity scores.
Comparison between coronary angiography and SPECT-MPI
All MPI studies exhibiting stress perfusion defects were considered abnormal. The overall level of disagreement (number of falsenegative and false-positive cases) for CAD detection against coronary angiography results was comparable in Flash 3D and in FBP imagebased reconstructions, (4 cases vs.5 cases, P=NS). The distribution of true positive (TP), false positive (FP), true negative (TN), or false negative (FN) perfusion defects according to the reconstruction technique and anatomic localization of the stenotic lesions is shown in Table 3. The disagreements for the absence or presence of perfusion defects in the coronary artery territories using Flash 3D were slightly higher in the LAD and lower in the RCA coronary territory as compared to the FBP, but no statistical significance was reached.
Flash 3D | FBP | |||||||
---|---|---|---|---|---|---|---|---|
Overall | LAD | LCx | RCA | Overall | LAD | LCx | RCA | |
TP n(%) | 15 (75) | 8 (40) | 6 (30) | 10 (50) | 14 (70) | 8 (40) | 5(25) | 6 (30) |
FP n(%) | 4 (20) | 9 (45) | 9 (45) | 5(25) | 4 (20) | 4 (20) | 8 (40) | 3 (15) |
TN n(%) | 1 (5) | 2 (10) | 4 (20) | 4 (20) | 1 (5) | 7 (35) | 6 (30) | 6 (30) |
FN n(%) | 0 | 1(5) | 1(5) | 1(5) | 1 (5) | 1(5) | 1(5) | 5(25) |
Table 3: Assessment of MPI interpretation using Flash 3D and FBP against coronary angiography as the standard of reference.
Image quality evaluation
Flash 3D reconstruction algorithm significantly increases the image contrast by 31% (P=0.002) during stress (from 30 to 61%, P=0.005) and at rest (from 11 to 42%, P< 0.001). A trend towards a decrease in SNR and CNR values was found during stress (from 9.04±3.34 to 7.64±2.68 for SNR, P=NS and from 0.52±0.84 to 0.22±0.11 for CNR, P=NS) and at rest (from 9.34±2.93 to 8.56±3.23 for SNR, P=NS and from 0.71±1.49 to 0.63±0.36 for CNR, P=NS) as compared with FBP.
The present study was performed to assess the feasibility of an OSEM reconstruction algorithm and CT- based attenuation correction (Flash 3D), which has been reported to improve image quality and quantitation performance by applying compensation techniques for photon attenuation, Compton scatter and 3-dimensional depthdependent resolution [12,13,15]. First, we compared the impact of Flash 3D versus FBP on myocardial perfusion defect and severity quantification in 20 patients who had undergone both stress-rest myocardial perfusion SPECT and coronary angiography. Our results indicated that there was no statistically significant difference in myocardial perfusion defect extent and severity using Flash 3D versus FBP. This study showed that the number of patient images with perfusion defects was identical with Flash 3D and FBP reconstruction. Small differences in perfusion defect extent (up to 6%) were encountered in individual coronary artery territories with Flash 3D against FBP, except the RCA vascular territory (17% mean difference during stress), but no statistical significance was reached. The difference in mean severity scores was negligible, with the highest value of 0.3 SD’s found in the LAD coronary territory during stress. With respect to individual vascular territories (Table 3), the disagreements between the two types of reconstruction were mostly in the RCA and LCx arteries. This is probably due to the effect of subdiaphragmatic activity and diaphragmatic attenuation especially in the case of inferior wall which is not corrected in FBP reconstruction. This finding may suggest that, although the compensation techniques for the photon attenuation and scatter might be effective with Flash 3D, it does not decrease the difficulties associated with the inferior wall attenuation artifacts [16]. Also, a certain degree of misregistration might be present between SPECT and CT data, even after manual realignment. Scatter as well, may contribute, since all patients underwent pharmacologic stress with dipyridamole, well-known for the high radiopharmaceutical uptake in the liver [17]. Nevertheless, the use of a standard template of coronary arteries distribution in MPI perfusion analysis may have also influenced. For instance, if a perfusion defect lies in the RCA territory, slightly increased perfusion defect size in Flash 3D reconstruction will incorporate part of the LCx territory as well [18]. Summarizing, Flash 3D algorithm provides results equivalent to conventional FBP in terms of quantitative myocardial perfusion defect analysis and location as confirmed by coronary angiography.
Like previous studies [19-21] our results indicate that combined scatter, attenuation and detector response correction led to an increase in image quality. The contrast between the myocardial wall and the left ventricular cavity increases by 31%. Without any corrections (FBP reconstruction), the CNR and SNR values were higher, with a trend toward decreased values in Flash 3D reconstructed images that is consistent with a slight increase in image noise. This could be related to the increased number of iterations and subsets used in our study (10 iterations, 8 subsets) which may have an additive effect on noise as it has been suggested by other researchers [22,23]. Anyhow, Flash 3D reconstructed images are improved quantitatively by increasing the absolute values of image contrasts, while CNR and SNR values remain approximately equal to those measured in the FBP reconstructed images.
The small number of subjects studied is a major limitation of the study and subsequent large scale clinical implementation of this novel image reconstruction algorithm requires a further larger patient study rigorously tested in standardized conditions.
OSEM 3D image reconstruction algorithm (Flash 3D) with attenuation, scatter, and spatial resolution compensation proved to be superior to conventional FBP in terms of image contrast, and equivalent to FBP for specific clinical findings, and in particular does not decrease the difficulties associated with the inferior wall attenuation artifacts.
The authors wish to thank the staff of the Centre for Nuclear Medicine/PET for their excellent contribution in imaging patients.