Andrzej Polanczy, Aleksandra Piechota-Polanczyk, Martin Funovics, Christoph Domenig, Josif Nanobashvili, Christoph Neumayer and Ihor Huk
Medical University Vienna, Austria
Lodz University of Technology, Poland
Jagiellonian University, Krakow, Poland
Scientific Tracks Abstracts: J Clin Exp Cardiolog
Introduction: The aim of the study was to investigate a mechanical behavior of various types of artificial vessels in relation to
the iliac arteries with the use of dedicated ex vivo bioengineering reactor.
Methods: Artificial circulatory model (ACM) for computational projection of vessel structure under different flow conditions
was designed and built. Analyzed vessels were supplying with the homemade fluid mimicking blood. Following types of vessels
were analyzed: synthetic prostheses (ePTFE, ePTFE with spiral, Dacron), biosynthetic prostheses (Omniflow II, gelatin sealed
ePTFE), iliac arteries and silicon tubes. Each time the same length of vessel (100 mm) was analyzed. Mechanical behavior was
introduced with the use of following parameters: change of diameter, wall displacement, deformation factor and divergent
factor. Moreover, ACM results were verified with medical data. 2D-speckle-tracking-technique (2DSTT) was applied to assess
diameter dilatation for patients with ePTFE prostheses/iliac arteries.
Results: It was presented that both Dacron and bio prostheses act similarly to real tissue. Approximately 21% difference
for gelatin sealed ePTFE prostheses and 25% for Dacron prostheses compare to iliac arteries was observed. While, ePTFE
prostheses presented about 2.4-fold increase of stiffness compare to the flexibility of iliac arteries. Moreover, ePTFE prostheses
act much more like silicon tubes rather than iliac arteries.
Conclusions: Artificial reconstruction of blood flow in different spatial configuration of human and artificial vessels allows
simulating different mechanical response of tissue vascular grafts and silicon tubes.
Recent Publications
1. Polanczyk A, Piechota-PolanczykA and Stefanczyk L (2017) A new approach for the pre-clinical optimization of a
spatial configuration of bifurcated endovascular prosthesis placed in abdominal aortic aneurysms. PLOS One DOI:
10.1371/journal.pone.0182717.
2. Polanczyk A, Podyma M, Trebinski L, Chrzastek J, Zbicinski I and Stefanczyk L A (2016) Novel attempt to standardize
results of CFD simulations basing on spatial configuration of aortic stent-grafts. PLOS One DOI: 10.1371/journal.
pone.0153332.
3. Polanczyk A, Podyma M, Stefanczyk L, Szubert W and Zbicinski I (2015) A 3D model of thrombus formation in a stentgraft
after implantation in the abdominal aorta. Journal of Biomechanics 48(3):425-431.
4. Mundargi R, Venkataraman D, Kumar S, Mogal V, Ortiz R and Loo J (2015) Novel sensor-enabled ex vivo bioreactor:
A new approach towards physiological parameters and porcine artery viability. BioMed Research International DOI:
10.1155/2015/958170.
5. F Prandi, M Piola, M Soncini, C Colussi, Y D'Alessandra, E Penza et al. (2015) Adventitial vessel growth and progenitor
cells activation in an ex vivo culture system mimicking human saphenous vein wall strain after coronary artery bypass
grafting. PLOS One 10:e0117409.